FISHER & PAYKEL HEALTHCARE LTD F&P OPTIFLOW JUNIOR 2; NASAL CANNULA Back to Search Results
Model Number OJR414
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Event or Problem Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that the tubing of two ojr414 optiflow junior 2 nasal cannula's had detached from the cannula end.There was no reported patient consequences.
 
Additional Manufacturer Narrative
(b)(4).Method: the subject devices, ojr414 optiflow junior 2 nasal cannula were returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is thus based on the evaluation of the returned devices, information provided by the distributor, and our knowledge of the product.Results: evaluation of the subject devices confirmed that one of the tubes was detached from the cannula end.Conclusion: based on the evaluation of the returned device, information provided by the distributor, and our knowledge of the product, the reported event could have resulted from the cannula being subjected to excessive force.As part of the manufacturing process, all optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails specifications is discarded.In addition, representative samples from each batch are functionally tested to assess retention strength between the assembled components.If there are any failures the entire batch is quarantined for further investigation.The user instructions illustrate in pictorial format the correct set-up and proper use of the optiflow junior 2 nasal cannula.They also state the following: - "appropriate patient monitoring (e.G.Oxygen saturation) must be used at all times.Failure to monitor the patient (e.G.In the event of an interruption to gas flow) may result in serious harm or death)." - "ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." - "do not wrap, insulate, stretch or crush the tubing as this may impair the performance of this product or compromise safety (including potentially causing patient harm).
 
Event or Problem Description
A distributor in japan reported on behalf of a healthcare facility via a fisher & paykel healthcare (f&p) field representative that the tubing of two ojr414 optiflow junior 2 nasal cannulas detached from the cannula end.There was no reported patient consequences.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
F&P OPTIFLOW JUNIOR 2
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key23774361
Report Number9611451-2025-01894
Device Sequence Number10057661
Product Code BTT
UDI-Device Identifier09420012435934
UDI-Public(01)09420012435934(10)2103806010(11)250611(17)300611
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K222197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/18/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOJR414
Device Catalogue NumberOJR414
Device Lot Number2103806010
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2025
Supplement Date Received by Manufacturer02/16/2026
Initial Report FDA Received Date12/10/2025
Supplement Report FDA Received Date03/17/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/11/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-