U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD FISHER AND PAYKEL HEALTHCARE; NASAL CANNULA Back to Search Results
Model Number OJR414
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/13/2025
Type of Reportable Event Malfunction
Event or Problem Description
A distributor in japan on behalf of a healthcare facility reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr410 optiflow junior 2 nasal cannula was detached from the cannula end.There was no reported patient consequence.
 
Additional Manufacturer Narrative
(b)(6).Fisher & paykel healthcare (f&p) is currently in the process of completing our investigation.We will provide a follow up report upon completion of our investigation.
 
Additional Manufacturer Narrative
(b)(6).Method: the subject device, ojr410 optiflow junior 2 nasal cannula was not returned to fisher & paykel healthcare (f&p) for evaluation.Our investigation is based on the analysis of the information provided by the healthcare facility, and our knowledge of the product.Results: a healthcare facility in japan reported that the tubing of an ojr410 optiflow junior 2 nasal cannula got detached from the cannula end.Conclusion: without the return of the subject device or photography for evaluation, we are unable to determine the cause of the reported event.As part of the manufacturing process, all optiflow junior cannulas are 100% leak and occlusion tested after final assembly and any cannula that fails specifications is discarded.In addition, representative samples from each batch are functionally tested to assess retention strength between the assembled components.If there are any failures the entire batch quarantined for further investigation.The user instructions which accompany the ojr410 optiflow junior 2 nasal cannula show in pictorial format the correct placement and fitting of the cannula, and provide the following warnings: - "appropriate patient monitoring (e.G., oxygen saturation) must be used at all times.Failure to monitor the patient may result in loss of therapy, serious harm or death." - "ensure all connections are secure during use.Check the cannula is undamaged and that the flow path is maintained.Under excessive load, the cannula may disconnect to prevent forces being transferred to the patient." - "do not stretch the cannula on application; this may cause increase pressure to the patient's skin.If necessary, the cannula may be repositioned.".
 
Event or Problem Description
A distributor in japan on behalf of a healthcare facility reported via a fisher & paykel healthcare (f&p) field representative that the tubing of an ojr410 optiflow junior 2 nasal cannula was detached from the cannula end.There was no reported patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER AND PAYKEL HEALTHCARE
Common Device Name
NASAL CANNULA
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
diego vargas banuelos
17400 laguna canyon road
suite 300
irvine, CA 92618
9494341005
MDR Report Key23774365
Report Number9611451-2025-01890
Device Sequence Number10057663
Product Code BTT
UDI-Device Identifier09420012435743
UDI-Public(01)09420012435743(10)2103394902(11)241010(17)291010
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K222197
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative,Distributor
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/29/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberOJR414
Device Catalogue NumberOJR414
Device Lot Number2103394902
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/15/2025
Supplement Date Received by Manufacturer01/06/2026
Initial Report FDA Received Date12/10/2025
Supplement Report FDA Received Date01/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured10/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-