MAUDE Adverse Event Report: MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS MAQUET POWERLED II; LAMP, SURGICAL Back to Search Results
Model Number ARDPWT239222A
Medical Device Problem Code Peeled/Delaminated (1454)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event site name: (b)(6).Additional information will be provided following the conclusion of the investigation.The unique identifier (udi) # information is not available as this medical device/model was marketed and/or discontinued in us before the udi requirement became mandatory.
 
Event or Problem Description
Getinge became aware of an issue with one of surgical lights - powerled ii.It was stated the paint is peeling off around the control panel.We decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure may cause contamination.
 
Additional Manufacturer Narrative
The correction of d1 brand name, d4 serial #, d4 catalog # and d4 version or model # fields deems required.This is based on the internal evaluation.Previous d1 brand name: powerled ii.Corrected d1 brand name: maquet powerled ii.Previous d4 serial #: (b)(6).Corrected d4 serial #: (b)(6).Previous d4 catalog #: ard569202911.Corrected d4 catalog #: n/a.Previous d4 version or model #: ard569202911.Corrected d4 version or model #: ardpwt239222a.Getinge became aware of an issue with one of surgical lights - powerled ii.It was stated the paint is peeling off around the control panel.We decided to report the issue in abundance of caution as any paint particles falling off into sterile field or during procedure may cause contamination.Based on the information collected, it was established that when the event occurred, the surgical light did not meet its specification and in this way the device contributed to event.It is unknown if claimed devices were or were not being used for patient treatment or diagnosis when the event took place.A root cause analysis for paint peeling problem was performed by subject matter experts and concluded that all maquet sas products comply with: current version of iec 60601-1 general requirements for basic safety and essential performance.Current version of iec 60601-2-41 particular requirements for the safety of surgical luminaires and luminaire for diagnosis.Paint definition pfc066.This procedure defines maquet sas¿s requirements for all painted parts.Disinfection products test: the aim of these tests is to detect any incompatibility with disinfectant.The most probable root cause of paint degradation would be an incorrect cleaning procedure/non-compatible cleaning product.Such troubles should have been detected first by the user during daily and monthly checks.The current cleaning product must be improved as described in the user manual.The damaged parts must be replaced as soon as possible.The paint chip or paint damages are due to: impacts, collisions (abnormal use).The operating manual includes the instructions to pre-position the arms prior to use, in order to prevent damages.To prevent any similar incident, it is recommended to avoid collisions between devices.Visual inspections during the cleaning allow to detect the painting defect, we recommend to perform corrective maintenance to rectify the default after its detection.Minor paint chip can be repaired with touch up paint, nevertheless, the parts impacted by serious damage must be replaced.Getinge shall continue to monitor for any further events of this nature and does not propose any further action at this time.
 
Event or Problem Description
Manufacturer's reference number (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAQUET POWERLED II
Common Device Name
LAMP, SURGICAL
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
sylviane boiron bourgeois
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key23775748
Report Number9710055-2025-0000305
Device Sequence Number10131825
Product Code FTD
UDI-Device Identifier03700712410216
UDI-Public(01)03700712410216
Combination Product (Y/N)N
Initial Reporter CountryGM
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Model NumberARDPWT239222A
Device Catalogue NumberARD569202911
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/27/2025
Supplement Date Received by Manufacturer02/26/2026
Initial Report FDA Received Date12/11/2025
Supplement Report FDA Received Date03/09/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/03/2020
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-