| Model Number |
CNA0T0 |
| Medical Device Problem Code |
Inaccurate Delivery (2339)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
11/05/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A pharmacist reported that, during intraocular lens (iol) implantation surgery, the mechanism became blocked, and the lens could not be implanted.The surgery was completed using different lens.Additional information was received stating that, the implant remained stuck in the corneal incision at the time of implantation and not in the injector.Since the surgeon had to remove it forcefully with tweezers, he decided to replace it with a new implant.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for analysis, and the reported complaint was observed.The device was returned loose in the product carton.The lock-out assembly has been removed.Viscoelastic is observed in the device.The plunger has been fully advanced outside the nozzle tip and is advanced under the intra ocular lens (iol).The iol is advanced into the mid nozzle with the leading haptic partially exiting the nozzle tip.We are unable to determine the root cause for the reported complaint "this lens could not be implanted because the mechanism became blocked".Plunger underride was observed.Plunger underride may occur: ¿ if the lens has become misaligned in the lens bay during the manufacturing process.¿ due to the use of a non-qualified viscoelastic.Material properties of non-qualified ophthalmic viscosurgical device (ovd)s may contribute to underfill, overfill, misfolding of the haptics, or other inconsistent folding outcomes.¿ if inadequate viscoelastic is placed in the device this will cause inadequate coverage of the lens fold path which may cause the lens to advance incorrectly or become ¿stuck¿ in the device allowing the plunger to underride the lens.¿ if the operating room temperature is too high (> 23 degree c / 73 degree f) lens folding consistency is negatively affected, the lens is more adherent, and this may inhibit lens advancement.Any of the above listed causes alone, or in combination, may create the reported event.Based on the results from the product history record, the products met release criteria.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|