| Catalog Number |
BOET510 |
| Medical Device Problem Codes |
Patient-Device Incompatibility (2682); Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Hypersensitivity/Allergic reaction (1907); Unspecified Infection (1930)
|
| Date of Event |
09/05/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
The doctor reports that the dental implants failed due to allergic reaction & infection.The doctor reports in the per that the procedure was not concluded by placing another implants.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).A1: patient identifier unknown / not provided.D10: concomitant medical product: boet410, t3® ex hex tapered implant 4 x 10mm, lot: 2022120758.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.D10.Concomitant medical product updated: oss485, osseotite® implant 4 x 8.5mm, lot: unknown.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H6: adverse event problem.H10: additional narrative.
|
| |
|
Event or Problem Description
|
|
No further event information is available at the time of this report.
|
| |
|
Additional Manufacturer Narrative
|
|
Zimvie complaint number (b)(4).The following fields have been updated: b4: date of this report.B5: describe event or problem.G3: date received by manufacturer.G6: type of report.H1: type of reportable event.H2: follow up type.H3: device evaluated by manufacturer.H6: adverse event problem.H11: additional narrative.Zimvie received one (1) boet510, (t3® ex hex tapered implant 5 x 10mm) for evaluation.Visual evaluation was performed, the implant had signs of use with debris on its external threads.There was no damage identified or signs of malfunction that would have contributed to the event.Measurements match drawing.Dhr review was completed for the subject lot number 2022020796.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number 2022020796 for similar events and no other complaint was identified.Review completed utilizing keywords: ¿dental: medical: allergic reaction, dental: medical: infection¿ based on the investigation and risk management file review, the most likely root cause determined from the investigation was patient factors.For infection: a summary investigation has been completed for infection events identifying that a definitive root cause cannot be identified due to the wide range of external (non-design/ non-manufacturing related) factors potentially impacting the sterility of the implant and its environment.Should additional information be received which indicates that the device may have caused or contributed to the infection event, an additional report will be submitted.Therefore, based on the available information, a device malfunction did not occur.No damage to the implant was identified.The reported event was non-verifiable.Infections is a medical condition and is non-verifiable with all the information.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the products were nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.At this time, the complaint investigation has been completed and the record will be closed.If additional information is received, the record will be re-opened for further evaluation.
|
| |
|
Search Alerts/Recalls
|