| Catalog Number |
119316M |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/03/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that they had another one of these foley trays today ref# (b)(4) that did not deflate.They believe this is lot# pk7676940, they were sent a picture of the sheet.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because this investigation did not result in any additional findings and no sample was available for evaluation.The reported event is addressed within the labeling as it states, to deflate catheter balloon: gently insert a luer lock or slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate to deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed, if this fails contact adequately trained professional for assistance, as directed by hospital protocol.A dhr could not be performed since no lot number was provided.No additional actions can be taken at this time.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that they had another one of these foley trays today ref# (b)(4) that did not deflate.They believe this is lot# pk7676940, they were sent a picture of the sheet.
|
| |
|