It was reported that during a cruciate ligament reconstruction surgical procedure, when using the cortical fxation fxd loop 25mm suture device, it was observed that the suture broke off when it was tightened.It was reported that another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.
|
|
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H11 additional narrative: investigation summary the product was not returned to depuy synthes, however a photo was provided for evaluation.Visual inspection of the returned photo found that the device is shown in used condition.The white suture was found broken and frayed.An investigation was made with the manufacturer; as a result, the white suture is utilized to assembly onto a graft construction which provides fixation.The green/white suture works as the lead suture for pulling the implant/graft construction with tension, the tension values are associated with this final implant assembly; the strength requirement for the sutures are 250n clinical spec ~1700n.The clinical spec tension is high compared with the expected intraoperative loads of 20-50n.Therefore, based on the work instruction of finish goods, the tension is measured only on the green/white suture during the manufacturing process.Since the white suture is broken and considering the damaged condition, a pull test is not required.Regarding the white suture, an inspection is performed during the manufacturing process to verify its size and condition, this information corresponds with the process control, and its record guarantees the white suture inspection.Since the suture breakage couldn't be originated during manufacturing process, the breakage can be attributed to procedure variables such as excessive counter tension or other instrument that pulls the white suture which creates a stress point on the suture, leading to a breakage.As per the instructions for use (ifu) the steps for a proper insertion are provided.The overall complaint was confirmed as the observed condition of the cortical fxation fxd loop 25mm would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to procedure variables such as excessive counter tension or other instrument that pulls the white suture which creates a stress point on the suture, leading to a breakage.As per ifu the steps for a proper insertion are provided.It has been determined that no corrective and/or preventative action is required.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
|