MAUDE Adverse Event Report: DEPUY IRELAND - 3015516266 ARTICULEZE HD CER 40 +8.5; PROSTHESIS, HIP, SEMI-CONSTR, METAL/CERAMIC/POLYM, CEMENT OR NONPOROUS, UNCEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 3015516266 ARTICULEZE HD CER 40 +8.5; PROSTHESIS, HIP, SEMI-CONSTR, METAL/CERAMIC/POLYM, CEMENT OR NONPOROUS, UNCEMENT Back to Search Results
Catalog Number 473440085
Medical Device Problem Codes Improper or Incorrect Procedure or Method (2017); Noise, Audible (3273)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2025
Type of Reportable Event Serious Injury
Event or Problem Description
Patient was reporting "clanking" during post op pt.X-rays looked in order according to the surgeon, so their assumption was that the liner may not have engaged fully.Patient was taken back to surgery; stem and femoral head were explanted.The cup and liner were solid with no visible issue according to the surgeon.The surgeon determined the stem to be undersized and loose at non-cemented interface.Femur was rebroached up to a size 7 actis collared; had good axial and rotational stability.A +12mm ts head was impacted and the leg was reduced.Construct was stable, had good schuck, and was not dislocating/impinging.Doi: (b)(6) 2025.Dor: (b)(6) 2025.Affected side: left hip.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional narrative: d4: udi: as the lot number for the device involved in the event was not provided, the full udi is currently not available.H3, h6: the device lot number is unknown; therefore, a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.No device associated with this report was received for examination.An evaluation of the manufacturing record could not be performed as the required product/lot number was not provided to complete the evaluation.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTICULEZE HD CER 40 +8.5
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTR, METAL/CERAMIC/POLYM, CEMENT OR NONPOROUS, UNCEMENT
Manufacturer (Section D)
DEPUY IRELAND - 3015516266
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
JTE WARSAW MFG SITE
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
elba bello
700 orthopaedic dr.
warsaw, IN 46581
9083863534
MDR Report Key23778902
Report Number1818910-2025-21666
Device Sequence Number15973469
Product Code LZO
UDI-Device Identifier10603295505778
UDI-Public(01)10603295505778
Combination Product (Y/N)N
Initial Reporter StateNM
Initial Reporter CountryUS
PMA/510(K) Number
K222296
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number473440085
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/20/2025
Initial Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP ACETABULAR LINERS
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
Patient SexMale
-
-