MAUDE Adverse Event Report: ETHICON INC. VCL+ UD 27IN 3-0 S/A CP-2; SUTURE, ABSORBABLE, SYNTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. VCL+ UD 27IN 3-0 S/A CP-2; SUTURE, ABSORBABLE, SYNTHETIC Back to Search Results
Catalog Number VCP868H
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Date sent to the fda: 12/11/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One winding former with one detached needle and one suture outside the foil packet that pertains to product code vcp868 was received for analysis.During visual inspection of the detached needle, the swage and attachment area was noted to be as expected.The barrel hole was examined, and remnant of suture was observed.The suture was examined, and it was found that it has begun the degradation process which caused the breakage suture.The foil was visually inspected, and excessive wrinkles and holes in the cavity were observed.The reported event is confirmed.The condition of the package is indicative of handling and storage issues that occurred outside of ethicon control.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
 
Event or Problem Description
It was reported that a patient underwent an unknown procedure on an unknown date and suture was used.During an unknown surgery, the suture was broken when the package was opened.It was not a control release needle.Another product was used to complete the case.There were no adverse consequences to the patient.Additional information was requested.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL+ UD 27IN 3-0 S/A CP-2
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-SAN ANGELO
3348 pulliam st
san angelo TX 76905
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key23779242
Report Number2210968-2025-13896
Device Sequence Number10083995
Product Code GAM
UDI-Device Identifier10705031055896
UDI-Public(01)10705031055896(17)260531(10)RG2AQB
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K032420
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/31/2026
Device Catalogue NumberVCP868H
Device Lot NumberRG2AQB
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2025
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/14/2025
Initial Report FDA Received Date12/11/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/24/2021
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-