Product complaint # (b)(4).Date sent to the fda: 12/11/2025.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.H3 investigational summary: the product was returned to ethicon for evaluation.One winding former with one detached needle and one suture outside the foil packet that pertains to product code vcp868 was received for analysis.During visual inspection of the detached needle, the swage and attachment area was noted to be as expected.The barrel hole was examined, and remnant of suture was observed.The suture was examined, and it was found that it has begun the degradation process which caused the breakage suture.The foil was visually inspected, and excessive wrinkles and holes in the cavity were observed.The reported event is confirmed.The condition of the package is indicative of handling and storage issues that occurred outside of ethicon control.As part of the ethicon quality process, all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|