The device involved in this event was discarded; therefore, no visual or physical analysis of the product could be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing and packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted in the device instructions for use, "free movement of the guidewire within the catheter is an important feature of a steerable guidewire system because it gives the user valuable tactile information.Test the system for any resistance prior to use." at this time, it is not possible to assign a definitive root cause for the event as reported.Based on the information provided, it appeared that clinical and/or procedural factors have contributed to the event as reported.We reviewed the associated risk documentation relative to this product and confirmed that this incident is listed and that it is trending within the established occurrence threshold.Lake region's effectiveness of design verification activities brings the overall risk down to as far as possible.Patient information was not provided; therefore, part a could not be completed.The sections left blank or unknown on this form could not be completed due to missing information.Attempts to gather further details to obtain the information were made and no additional information was provided regarding this event at the time of this report.Should additional information be received, a follow up medwatch report will be submitted.Note: although annex g instructions state that with stand-alone devices annex g term should not be used, code 4755 was selected due to component code (g) showing up on the missing data report.
|