| Model Number |
Ø8X 30MM BC IF SCRW, VENTED |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/21/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during an arthroscopic surgery the implant broke while it was screwed into the bone.The broken pieces were retrieved out of the patient.Per complaint reporter there was no harm for patient, operator or third party.The surgery was finished successfully with a new device with the same part number.It was not necessary to switch the surgical technique or do a second surgery.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation is in process.A follow-up report will be provided upon availability of additional information.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d9, g3, h3, h6.-complaint allegation is confirmed.-visual inspection of the fastthread¿ biocomposite¿ interference screw 8 x 30 mm, identified that the tip was broken off and the threads were damaged.-functional testing was not performed due to the damage to the device.The method of bone preparation and the quality of the bone encountered were not specified.Per dfu-0111-eo - g.Precautions - 4.Bio-tenodesis screw only: use the appropriate size arthrex drill to create a pilot hole in the bone.6.Bio-absorbable interference screw only: it is important to completely seat the screwdriver to prevent potential stripping of the hex and/or screw fracture during insertion or removal.The most likely cause for the reported failure can be attributed to user error of the device due to improper bone preparation, misalignment insertion, and/or prying/leveraging of the device during insertion.
|
| |
|
Search Alerts/Recalls
|