BOSTON SCIENTIFIC CORPORATION SENTINEL; TEMPORARY CATHETER, EMBOLIC PROTECTION, TRANSCATHETER INTRACARDIAC PROCEDURES Back to Search Results
Model Number CMS15-10C-US
Medical Device Problem Codes Nonstandard Device (1420); Difficult to Remove (1528); Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/08/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that failure to capture filter occurred.A transcatheter aortic valve replacement (tavr) procedure was being performed with a sentinel cerebral protection system (cps) for embolic protection.The sentinel cps was advanced to the intended location.The distal filter was deployed and recaptured three times; however, difficulty was encountered when deploying the distal filter and the user experienced resistance with the distal filter slider (#3).During withdrawal of the sentinel cps, the proximal filter was fully retracted, however approximately 1cm of the distal filter, near the distal tip, could not be completely retracted.The sentinel cps was removed in this state with no issues.The physician commented that the distal tip was unusual, suggesting a possible defect.The physician also commented that the patient's anatomy contained mild stenosis at the origin of the left common carotid artery which may have contributed to the issue.A new sentinel cps was utilized to complete the procedure.No adverse events occurred.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENTINEL
Common Device Name
TEMPORARY CATHETER, EMBOLIC PROTECTION, TRANSCATHETER INTRACARDIAC PROCEDURES
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
2546 calle primera
alajuela
CS  
Manufacturer Contact
mark mischke
4100 hamline ave n
arden hills, MN 55112
7634941632
MDR Report Key23782371
Report Number2124215-2025-90447
Device Sequence Number21189387
Product Code PUM
UDI-Device Identifier00863229000004
UDI-Public(01)00863229000004(17)270722(10)37028182
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K192460
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative,Distributor
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberCMS15-10C-US
Device Catalogue NumberCMS15-10C-US
Device Lot Number0037028182
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/08/2025
Initial Report FDA Received Date12/11/2025
Date Device Manufactured07/22/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age84 YR
Patient SexFemale
-
-