• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN SURGICAL TRI-STAPLE 2.0; STAPLE, IMPLANTABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN SURGICAL TRI-STAPLE 2.0; STAPLE, IMPLANTABLE Back to Search Results
Model Number SIG45AXT
Medical Device Problem Code Failure to Form Staple (2579)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
Date of Event 11/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during a right vats (video-assisted thoracic surgery) wedge resection of the lower and middle lobes, the first reload experienced incomplete deployment of staples during the final firing across lung tissue containing a pulmonary artery branch, resulting in only a few visible completed staples at the apex, with the remaining staples either not found or bunched at the distal end and not completely formed.This led to bleeding from a pulmonary artery branch, requiring the immediate closure of jaws to apply pressure to the tissue and vessel, calling blood into the room, opening all thoracotomy trays in preparation for possible conversion to open surgery, and bringing in a second stapler setup.The procedure was extended by 30 minutes.The team was able to clamp off the section and fire another reload to complete the wedge resection without additional issue.The second reload when fired across the parenchyma, experienced a break at the articulation joint, with the broken part disengaging but not falling into the patient cavity, and the blowout plate was exposed and ribboned out at the wrist joint during retraction of the knife blade while in a fully articulated position.No x-ray was needed to locate components in the patient cavity.The handle, adapter and reload were replaced after the second reload (black 45) was fired.It was noted that the same adapter and handle were used for both the purple and black 45 firings.
 
Additional Manufacturer Narrative
D10 concomitant products: egia45amt, egia45amt egia 45 artic med thick sulu, (lot# p5g1173) sigadaptstnd, sig power sigadaptstnd linear adapter (serial# (b)(6)) sigphandle, sig power sigphandle handle (serial# (b)(6)), medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Additional Manufacturer Narrative
Additional information: g3, h3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device and a photo were available for evaluation.A video was also provided.Visual inspection noted that the reload was fully fired with the interlock engaged.The reload had damage to the cutting edge of the knife blade.Pushers were noted to be sub-flush along the majority of the cartridge.Functional testing found that the reload was loaded into a representative powered handle running v29.6 software.The reload was noted to have been marked as used.The reload was then loaded into a representative manual handle.The interlock was overridden and the reload was applied to test media.Test media was transected.The interlock was tested and found to function properly.It was reported that the staple line was incomplete and there was poor staple formation.The reported issues were confirmed.The product analysis noted evidence that the device was not used as intended.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.The instructions included with this device provide the following guidance: when positioning the stapler on the application site, ensure that no obstructions, such as clips, are incorporated into the instrument jaws.Firing over an obstruction may result in incomplete cutting action and/or improperly formed staples.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRI-STAPLE 2.0
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23782460
Report Number1219930-2025-05683
Device Sequence Number14699878
Product Code GDW
UDI-Device Identifier10884521543607
UDI-Public(01)10884521543607(17)300531(10)N5F1650Y(20)02
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K160176
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/10/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSIG45AXT
Device Catalogue NumberSIG45AXT
Device Lot NumberN5F1650Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2025
Supplement Date Received by Manufacturer02/03/2026
Initial Report FDA Received Date12/11/2025
Supplement Report FDA Received Date02/10/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured06/05/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
SEE H11.
Patient Age58 YR
Patient SexMale
Patient Weight118 KG
-
-