• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MENTOR TEXAS MENTOR MEMORYGEL XTRA BREAST IMPLANT; PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED Back to Search Results
Catalog Number SMPX370
Medical Device Problem Codes Adverse Event Without Identified Device or Use Problem (2993); Migration (4003)
Health Effect - Clinical Code Capsular Contracture (1761)
Date of Event 11/18/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
On 21-jan-2026, mentor received additional information about the event: it was reported that the patient¿s right breast was painful, tight, and distorted.The patient¿s explanted right breast prosthesis was replaced with a mentor memorygel xtra breast implant gel breast prosthesis.Manufacturer¿s reference number: (b)(4).
 
Event or Problem Description
It was reported that a patient who underwent a breast augmentation revision procedure with a mentor memorygel xtra breast implant 370cc gel breast prosthesis developed baker grade iv capsular contracture in her right breast post implantation.The capsular contracture was diagnosed via a physical exam.As a result, the patient is scheduled to undergo explantation on (b)(6) 2026.
 
Additional Manufacturer Narrative
Since the device has not been returned for analysis, no product failure analysis can be conducted, and no determination of possible contributing factors can be made.As such, the investigation will be closed.If the complaint device is received in the future, the investigation will be reopened and conducted as appropriate.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: capsular contracture.D6b explantation date: (b)(6) 2026.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 06-mar-2026, the mentor failure analysis lab received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.Manufacturer¿s reference number: (b)(4).
 
Additional Manufacturer Narrative
On 31-mar-2026, mentor completed its investigation on the suspect medical device.Device evaluation summary: the product was returned to mentor for evaluation.Mentor conducted a visual inspection of the returned device.During visual evaluation, no apparent damage or visual anomalies were observed on the returned device.Regarding the reported capsular contracture condition, there are not enough details to determine the factors that may have caused this condition.Capsular contracture in the patient¿s breast might be the result of the body¿s individual physiological response to the implantation of a foreign object in soft tissue.Further, capsular contracture is a known complication associated with these devices and is referenced in our current product insert data sheet.Implant displacement/migration may occur from improper implant sizing and/or placement, i.E., when the implant is too large or the pocket too small or when there has been inadequate preoperative assessment of stresses causing movement of the prosthesis.Implant displacement/migration is a known complication associated with these devices and is referenced in our current product insert data sheet.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.Based on the results of this investigation, it was determined that the product met the manufacturing release criteria: since no malfunction was observed during the investigation.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR MEMORYGEL XTRA BREAST IMPLANT
Common Device Name
PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer (Section G)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
Manufacturer Contact
henry boden
3041 skyway circle north
irving, TX 75038
9497882078
MDR Report Key23782692
Report Number1645337-2025-13790
Device Sequence Number12044463
Product Code FTR
UDI-Device Identifier00081317025054
UDI-Public(01)00081317025054(17)281115(10)9982389
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P030053
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2024
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue NumberSMPX370
Device Lot Number9982389
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/18/2025
Supplement Date Received by Manufacturer01/21/2026
03/06/2026
03/31/2026
Initial Report FDA Received Date12/11/2025
Supplement Report FDA Received Date02/04/2026
03/30/2026
04/06/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured11/30/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age74 YR
Patient SexFemale
Patient EthnicityNon Hispanic
Patient RaceWhite
-
-