Model Number FX384T
Medical Device Problem Code No Flow (2991)
Health Effect - Clinical Code Hydrocephalus (3272)
Date of Event 10/20/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The adverse event is filed under aic reference (b)(4).Investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported that a progav 2.0 shunt system (#fx384t) was implanted during a procedure.According to the complainant, the shunt system is clogged.The patient underwent a revision procedure.The complained product will be sent to the manufacturer for investigation.No patient complications were reported as a result of the revision procedure.
 
Additional Manufacturer Narrative
The adverse event is filed under aic reference (b)(4).Additional information: d4 item information d6b expiration date d9 device returned to manufacturer date f9 approximate age of device.Visual inspection: during the investigation, no significant deformations or damage of the valve determined.Permeability test: a permeability test has shown that all components are permeable.Computer controlled test: to investigate the claim of under-drainage, the opening pressure is measured using a miethke computer controlled testing apparatus which simulates a cerebrospinal fluid flow.The valves are tested in both the horizontal as well as the vertical positions.The results show that the progav 2.0 operates within the accepted tolerance in the horizontal position.The shuntassistant 2.0 does not operate within the specified tolerances in the vertical position.An accelerated outflow of shuntassistant 2.0 could be determined.Adjustment test: the progav 2.0 was tested and is adjustable to all specified pressures.Braking force and brake function test: the brake functionality test has shown that the brake function operates as expected and the braking force is within the given tolerances.Results: please note that a comprehensive examination of the progav 2.0 shunt system is not possible because we received no approval for the destructive examination despite several requests.According to the results of the investigation, we cannot determine functional deviations of the progav 2.0.The valve operated within all specifications.Furthermore, an accelerated outflow of the shuntassistant 2.0 was detected.Regarding similar complaints where we could examine the inside of the products, we could determine that the most frequent cause for a deterioration in the function of the product are deposits caused by natural substances in the cerebrospinal fluid, e.G.Protein, blood or tissue particles.These are among the known and unavoidable risks and side effects of hydrocephalus therapy.Even small amounts of organic material can affect the integrity of the valve if aggregated in the valve.A final clarification of the cause what has led to accelerated outflow is only possible by an examination we can exclude a defect at the time of release.The shunt system met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.No further regulatory actions are required from our point of view.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH SA2.0 15 AND SPRUNG RESERVOIR
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key23782976
Report Number2916714-2025-00076
Device Sequence Number17413620
Product Code JXG
UDI-Device Identifier04041906502454
UDI-Public(01)04041906502454(17)280316(10)20064736
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Importer
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/26/2026
Report Date (Section F) 12/11/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberFX384T
Device Catalogue NumberFX384T
Device Lot Number20064736
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date11/18/2025
Device Age34 MO
Event Location Hospital
Date Report to Manufacturer11/18/2025
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/11/2025
Supplement Report FDA Received Date01/26/2026
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
Date Report Sent to FDA12/11/2025
-
-