It was reported that a 51-year-old caucasian, female patient underwent breast surgery with 275cc mentor smooth round spectrum on both sides and experienced breast implant deflation on her right side postoperatively.As a result, the device was explanted on (b)(6) 2025.According to the information received, the patient experienced deflation of the right-side implant.No product quality issues were reported for the left-side device.On december 5, 2024, the mentor failure analysis lab received two devices for evaluation and were both found damaged upon evaluation.Hence, this report details investigation results for the second device received as no clarification was received after multiple follow-ups.
|
|
Device evaluation summary: mentor conducted visual inspection and microscopic examination of the returned device.Visual analysis of the returned device determined that a tear was found on the posterior view of the breast implant, measuring approximately 2.5 cm.Microscopic examination was performed on the edges of the rupture, and parallel striations were found in the whole area of the tear.Parallel striations are consistent with markings made by a sharp object perforating the implant shell.Based on the information currently available, microscopic examination of the returned product indicates that the implant could have been damaged during or subsequent to implantation.The product insert data sheet cautions to not allow cautery devices or sharp instruments, such as scalpels, suture needles, hypodermic needles, hemostats, adson forceps or scissors to contact the device during the implantation or other surgical procedures.As part of mentor's quality process, all devices are manufactured, inspected, and released according to approved specifications.No corrective and preventive action (capa) is required now.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.Reason for device explant and/or reoperation: na.D4: udi: as the lot, and serial number for the device involved in the event was not provided, the full udi is currently not available.Mentor is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which mentor has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, mentor, or its employees that the report constitutes an admission that the device, mentor, or its employees caused or contributed to the potential event described in this report.If certain information is unknown, not available or does not apply, the section/field of the form is left blank.Manufacturer¿s reference number: (b)(4).
|