• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES PR FOGARTY; CATHETER, EMBOLECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES PR FOGARTY; CATHETER, EMBOLECTOMY Back to Search Results
Model Number 12TLW805F35
Medical Device Problem Code Failure to Deflate (4060)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/04/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device evaluation is anticipated.A supplemental report will be forthcoming when the investigation is completed.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
 
Event or Problem Description
It was reported that before use of the fogarty catheter, when bench testing, it would either not inflate or was difficult to inflate and then not deflate.There was no patient injury.
 
Additional Manufacturer Narrative
The device was not returned for product evaluation.Without return of the unit, it is not possible to determine if there was damage or a defect that existed on the unit that could have contributed to the event or if any procedural factors may have contributed to the reported issue.The device history record review was completed and the product passed without nonconformances.An engineering evaluation was initiated to assess for any manufacturing related processes which could be correlated to the complaint.As the complaint could not be confirmed, there is not sufficient evidence to determine a root cause.As part of the manufacturing process 100% of the units go through a balloon winding and full visual inspection.The instructions for use also contains the following warning : use of liquid inflation media will result in slower inflation/deflation rates.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FOGARTY
Common Device Name
CATHETER, EMBOLECTOMY
Manufacturer (Section D)
EDWARDS LIFESCIENCES PR
state rd indus pk 402 km 1.4
anasco
Manufacturer (Section G)
BD APM
17200 laguna canyon rd
irvine CA
Manufacturer Contact
erika bonilla
1 edwards way
irvine, CA 92614
MDR Report Key23785827
Report Number2015691-2025-09960
Device Sequence Number10442797
Product Code DXE
UDI-Device Identifier00690103043532
UDI-Public(01)00690103043532(17)270828(11)250529(10)66621569
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K240078
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/21/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number12TLW805F35
Device Lot Number66621569
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/26/2025
Supplement Date Received by Manufacturer01/20/2026
Initial Report FDA Received Date12/11/2025
Supplement Report FDA Received Date01/21/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured05/29/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-