|
Product event summary: the device was returned and analyzed.Analysis of the returned device was inconclusive.Analysis of the hybrid revealed the device set eri/rrt while implanted.Review of save to disk data showed that there was an unexplained voltage drop beginning on (b)(6) 2024.Both loaded and unloaded pacing current drain levels were normal for this device over the range of battery voltage it operated under during its service time.There was no evidence of a high current drain condition on the hybrid and no hybrid anomalies were observed.Analysis of the battery revealed the manufacturing data, the data gathered during the visual and dimensional inspection, and the electrical testing data, did not show any abnormal results.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|