| Model Number |
G22633 |
| Medical Device Problem Code |
Failure to Conduct (1114)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Additional information provided states this device was used with olympus and conmed active cords, which is incompatible.The ifu states "this device must only be used with an active cord compatible with a 3 mm diameter connector.This device has only been verified to be compatible with the following cook active cords: acu-1 and acu-1-vl (supplied non-sterile)." the instructions for use contain the following information to assist with proper set-up and use of the device: "before using this device, follow recommendations provided by electrosurgical unit manufacturer to ensure patient safety through proper placement and utilization of patient return electrode.Ensure a proper path from patient return electrode to electrosurgical unit is maintained throughout the procedure." "inspect active cord.Cord must be free of kinks, bends, breaks and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." "with electrosurgical unit off, prepare equipment.Securely connect active cord to device handle and electrosurgical unit.Active cord fittings should fit snugly into both device handle and electrosurgical unit.Following instructions from electrosurgical unit manufacturer, position patient return electrode and connect it to electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify desired settings and activate electrosurgical unit.Note: maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.Additional comments: based on the information provided that the device was used with an incompatible active cord, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
| |
|
Event or Problem Description
|
|
During a polypectomy in the colon, the physician used two cook acusnare polypectomy snares.It was reported that the snares were unable to remove the polyp tissue [when electrosurgical generator is activated, no electrosurgical current present at snare head].They used another device to finish the procedure.A section of the device did not remain inside the patient¿s body.The patient did not require any additional procedures due to this occurrence.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.However, 3 devices from the lot number provided in the report were returned and evaluated.Additionally, 8 devices from lot w4973840, 4 devices from lot w4978197, 5 devices from lot w4964401, and 1 device from lot w4964393 were returned and evaluated.Sealed devices #1-21: our laboratory evaluation of the sealed devices could not confirm the report as it was described, the devices functioned as intended.The devices were submitted to the engineering test lab (etl) and were functionally tested to verify that a polypectomy could be performed successfully.The etl determined that the devices met the requirements.The snares attached to the active cord, advanced through the scope, advanced out of the device, ensnared and electrosurgically incised tissue, retracted back into the device, and removed from the scope all as expected.Therefore, the complaint could not be confirmed for the sealed devices.The device history record for the lot number said to be involved and the lot numbers for the sealed returned devices were reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: used devices: we could not conduct a complete investigation because the product said to be involved was not returned for evaluation.A definitive cause for the reported observation could not be determined.Sealed devices: a definitive cause for the reported observation could not be determined because the products said to be involved functioned as intended.A discrepancy or anomaly that could have contributed to the reported observation was not observed during our laboratory analysis of the returned product.Additional information provided states this device was used with olympus and conmed active cords, which is incompatible.The ifu states "this device must only be used with an active cord compatible with a 3 mm diameter connector.This device has only been verified to be compatible with the following cook active cords: acu-1 and acu-1-vl (supplied non-sterile)." the instructions for use contain the following information to assist with proper set-up and use of the device: "before using this device, follow recommendations provided by electrosurgical unit manufacturer to ensure patient safety through proper placement and utilization of patient return electrode.Ensure a proper path from patient return electrode to electrosurgical unit is maintained throughout the procedure." "inspect active cord.Cord must be free of kinks, bends, breaks and exposed wires to allow for accurate transfer of current.If an abnormality is noted, do not use active cord." "with electrosurgical unit off, prepare equipment.Securely connect active cord to device handle and electrosurgical unit.Active cord fittings should fit snugly into both device handle and electrosurgical unit.Following instructions from electrosurgical unit manufacturer, position patient return electrode and connect it to electrosurgical unit." "following electrosurgical unit manufacturer's instructions for settings, verify desired settings and activate electrosurgical unit.Note: maximum rated input voltage for this device is 2kvp-p for cut mode and 5 kvp-p for coagulation mode." prior to distribution, all acusnare polypectomy snares are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: based on the quality engineering risk assessment no corrective action is warranted at this time.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.Additional comments: based on the information provided that the device was used with an incompatible active cord, a cook representative has been directed to contact the medical facility involved in an effort to promote further education and understanding related to appropriate usage of this product.
|
| |
|
Additional Manufacturer Narrative
|
|
This investigation is on-going.A follow-up emdr will be provided within 30 days of receipt of new information.
|
| |
|
Search Alerts/Recalls
|
|