MAUDE Adverse Event Report: WILSON-COOK MEDICAL INC (COOK ENDOSCOPY) CYTOMAX II DOUBLE LUMEN CYTOLOGY BRUSH; ENDOCSOCPIC CYTOLOGY BRUSH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILSON-COOK MEDICAL INC (COOK ENDOSCOPY) CYTOMAX II DOUBLE LUMEN CYTOLOGY BRUSH; ENDOCSOCPIC CYTOLOGY BRUSH Back to Search Results
Model Number G22674
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Code Foreign Body In Patient (2687)
Date of Event 11/18/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Continued: e1: country: hong kong.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a corrective action (capa) has been initiated to reduce occurrences of brush tip/coil spring detachment for cytology brush devices.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all cytomax ii double lumen cytology brushes are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action (capa) was initiated to further investigate device failure due to brush tip/coil spring detachment.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography [ercp], the physician used a cook cytomax ii double lumen cytology brush.It was reported [that] the doctor used brush to collect cells from the hepatic duct on (b)(6) 2024.The procedure was finished and patient went through ct scan afterwards [on (b)(6) 2025].From the ct scan report, doctor found out there might be some foreign body inside patient and hence they did another check up and found a broken filiform tip inside patient's duct.A section of the device did not remain inside the patient¿s body.The tip was found and retrieved in recent check up on (b)(6) 2025.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Event or Problem Description
During an endoscopic retrograde cholangiopancreatography [ercp], the physician used a cook cytomax ii double lumen cytology brush.It was reported [that] the doctor used brush to collect cells from the hepatic duct on (b)(6) 2024.The procedure was finished and patient went through ct scan afterwards [(b)(6) 2025].From the ct scan report, doctor found out there might be some foreign body inside patient and hence they did another check up, and found a broken filiform tip inside patient's duct.On 29dec2026, a photo was provided showing the detached portion of the brush tip/coil spring.A section of the device did not remain inside the patient¿s body.The tip was found and retrieved in recent checkup on (b)(6) 2025.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
 
Additional Manufacturer Narrative
Investigation evaluation: a product evaluation was performed only by the picture provided in response to this report because the product said to be involved was not provided to cook for evaluation.Without the product or substantial evidence to contradict the complaint, it is considered confirmed based solely on the photo provided and statements describing the event.The photo shows the detached portion of the brush tip/coil spring.The tip is also kinked/bent in the middle portion, and the proximal portion is stretched out.A photo of the lot number was not provided.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the photo provided confirmed the report.A corrective action (capa) has been initiated to reduce occurrences of brush tip/coil spring detachment for cytology brush devices.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all cytomax ii double lumen cytology brushes are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action (capa) was initiated to further investigate device failure due to brush tip/coil spring detachment.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CYTOMAX II DOUBLE LUMEN CYTOLOGY BRUSH
Common Device Name
ENDOCSOCPIC CYTOLOGY BRUSH
Manufacturer (Section D)
WILSON-COOK MEDICAL INC (COOK ENDOSCOPY)
4900 bethania station rd
winston salem NC 27105
Manufacturer (Section G)
WILSON-COOK MEDICAL INC (COOK ENDOSCOPY)
4900 bethania station rd
winston salem NC 27105
Manufacturer Contact
sabrina o'brien
4900 bethania station rd
winston salem, NC 27105
3367440157
MDR Report Key23789560
Report Number1037905-2025-00829
Device Sequence Number16065674
Product Code FDX
UDI-Device Identifier00827002226746
UDI-Public(01)00827002226746(17)261226(10)W4795731
Combination Product (Y/N)N
PMA/510(K) Number
K192908
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 01/23/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberG22674
Device Catalogue NumberDLB-35-1.5-S
Device Lot NumberW4795731
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/18/2025
Supplement Date Received by Manufacturer12/29/2025
Initial Report FDA Received Date12/12/2025
Supplement Report FDA Received Date01/23/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/26/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN ENDOSCOPE MAKE AND MODEL
Patient SexUnknown
-
-