| Model Number |
G22674 |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
11/18/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Continued: e1: country: hong kong.Investigation evaluation: a product evaluation was not performed in response to this report because the product said to be involved was not provided to cook for evaluation.The report could not be confirmed.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: a corrective action (capa) has been initiated to reduce occurrences of brush tip/coil spring detachment for cytology brush devices.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all cytomax ii double lumen cytology brushes are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action (capa) was initiated to further investigate device failure due to brush tip/coil spring detachment.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography [ercp], the physician used a cook cytomax ii double lumen cytology brush.It was reported [that] the doctor used brush to collect cells from the hepatic duct on (b)(6) 2024.The procedure was finished and patient went through ct scan afterwards [on (b)(6) 2025].From the ct scan report, doctor found out there might be some foreign body inside patient and hence they did another check up and found a broken filiform tip inside patient's duct.A section of the device did not remain inside the patient¿s body.The tip was found and retrieved in recent check up on (b)(6) 2025.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Event or Problem Description
|
|
During an endoscopic retrograde cholangiopancreatography [ercp], the physician used a cook cytomax ii double lumen cytology brush.It was reported [that] the doctor used brush to collect cells from the hepatic duct on (b)(6) 2024.The procedure was finished and patient went through ct scan afterwards [(b)(6) 2025].From the ct scan report, doctor found out there might be some foreign body inside patient and hence they did another check up, and found a broken filiform tip inside patient's duct.On 29dec2026, a photo was provided showing the detached portion of the brush tip/coil spring.A section of the device did not remain inside the patient¿s body.The tip was found and retrieved in recent checkup on (b)(6) 2025.According to the initial reporter, the patient did not experience any adverse effects due to this occurrence.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation evaluation: a product evaluation was performed only by the picture provided in response to this report because the product said to be involved was not provided to cook for evaluation.Without the product or substantial evidence to contradict the complaint, it is considered confirmed based solely on the photo provided and statements describing the event.The photo shows the detached portion of the brush tip/coil spring.The tip is also kinked/bent in the middle portion, and the proximal portion is stretched out.A photo of the lot number was not provided.The device history record for the lot number said to be involved was reviewed.A discrepancy or anomaly was not observed with the product that was released for distribution.Investigation conclusion: our evaluation of the photo provided confirmed the report.A corrective action (capa) has been initiated to reduce occurrences of brush tip/coil spring detachment for cytology brush devices.The product said to be involved is included in the scope of the corrective actions.Prior to distribution, all cytomax ii double lumen cytology brushes are subjected to a visual inspection and functional testing to ensure device integrity.A review of the device history record confirmed that the lot said to be involved met all manufacturing requirements prior to shipment.Corrective action: a corrective action (capa) was initiated to further investigate device failure due to brush tip/coil spring detachment.The product said to be involved is included in the scope of the corrective action.A review of the complaint history was conducted.Based on this review, the likelihood of this type of report is rare.Quality assurance will continue to monitor for complaint trends.
|
| |
|