| Model Number |
PCF-H180AI |
| Medical Device Problem Code |
Failure to Clean Adequately (4048)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/25/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was observed during the device evaluation, the air/water channel of the colonovideoscope had white residue.There were no reports of patient involvement.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to olympus for inspection.The device evaluation found the air/water channel of the colonovideoscope had white residue.Based on the results of the investigation, olympus confirmed the device had foreign material, but the type of material or root cause cannot be identified.A definitive root cause cannot be determined.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor the field performance of this device.
|
| |
|