DEPUY SYNTHES POWER TOOLS 6.5CM ADULT CRANI ATTACHMT; DRILL, SURGICAL, ENT (ELECTRIC OR PNEUMATIC) INCLUDING HANDPIECE
|
Back to Search Results |
|
| Catalog Number |
CRANI-A |
| Medical Device Problem Codes |
Break (1069); Device Damaged by Another Device (2915)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service and evaluation, it was determined that the neuro tip of the craniotome device was bent/damaged.It was noted in the service order that the bearings were damaged.This event did not occur during surgery.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2025.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|
Additional Manufacturer Narrative
|
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for this medwatch, a supplemental medwatch report will be filed as appropriate.H10 additional narrative: device evaluation: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the neuro tip of the device was bent/damaged, the device had bearing damage, and the cutter could not be inserted.It was noted that the dura guard was scratched, and the bar could not be inserted (bearing detached).It was further determined that the device failed pretest for visual assessment and cutter insertion assessment.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to wear.
|
| |
|
Search Alerts/Recalls
|
|
|