• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMVIE US CORP LLC TSV® BELLATEK® ENCODE® EMERGENCE HEALING ABUTMENT 3.5MM(D) 6.5MM(P) 5MM(H); DENTAL ABUTMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMVIE US CORP LLC TSV® BELLATEK® ENCODE® EMERGENCE HEALING ABUTMENT 3.5MM(D) 6.5MM(P) 5MM(H); DENTAL ABUTMENT Back to Search Results
Catalog Number TEEHA365
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zimvie complaint number (b)(4).Customer has indicated that the product is in process of being returned to zimvie for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported that the implant at tooth site #31 had a healing abutment with damaged threads.The general dentist had a hard time removing the healing abutment.The general dentist stripped the abutment and in doing so, the implant became loose.The implant was removed and replaced it with a new one.An additional appointment was anticipated for the tooth site to heal enough to perform a final torque test several months from the implant replacement date.
 
Additional Manufacturer Narrative
This report is being submitted to relay corrected data, additional information and device evaluation.Correction: d4 device lot information was submitted incorrectly as teeha365 on the initial report.The correct device lot number is 2638049.The following sections are being reported: b4: date of this report was updated.D4: device lot # was updated.G3: date received by manufacturer was updated.G6: type of report was updated.H2: type of follow up was updated.H3: device evaluated by manufacturer was updated.H6: type of investigation code was added: 4114, 4109, 3331 and 4111.H6: investigation findings code was added: 213.H6: investigation conclusions code was added: 4315.H10: narrative/data was updated.Zimvie did not receive the complaint device for evaluation.Visual inspection could not be performed.The investigation has been performed based on the available information using the item number, device history record (dhr) review, and risk management file.Functional testing to recreate the reported event could not be performed due to the nature of the device and event.This complaint refers to the specific device being investigated for this complaint record.Dhr review was completed for the subject lot number.No deviations or non-conformances, which could have caused or contributed to the reported event, were noted as part of the dhr.Complaint history review was performed for the reported lot number for similar events and no other complaint was identified.The customer did not submit images for the reported event.Based on the investigation and risk management file review, the most likely root cause determined from the investigation is incorrect techniques used - customer error.Therefore, based on the available information, a device malfunction could not be verified.Without device receipt, the reported event is non-verifiable.No further investigation or immediate capa / hhe/d escalation is required, as the complaint investigation did not confirm the product was nonconforming at the time of distribution, and no new failure mode, harm, or hazardous situation was identified through the investigation performed.
 
Event or Problem Description
No additional event information received at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TSV® BELLATEK® ENCODE® EMERGENCE HEALING ABUTMENT 3.5MM(D) 6.5MM(P) 5MM(H)
Common Device Name
DENTAL ABUTMENT
Manufacturer (Section D)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
ZIMVIE US CORP LLC
4555 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
mark  mashburn
4555 riverside drive
palm beach gardens, FL 33410
5619713230
MDR Report Key23794488
Report Number0001038806-2025-03820
Device Sequence Number5546546
Product Code NHA
UDI-Device Identifier00844868045793
UDI-Public(01)00844868045793(17)291029(10)2638049
Combination Product (Y/N)N
Initial Reporter StateMD
Initial Reporter CountryUS
PMA/510(K) Number
K212730
Exemption Number5645646
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2025
Device Explanted Year2025
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,User Facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 02/11/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Catalogue NumberTEEHA365
Device Lot Number2638049
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer Not provided
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date12/12/2025
Supplement Report FDA Received Date02/11/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured10/30/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Patient Age69 YR
Patient SexFemale
Patient EthnicityNon Hispanic
-
-