| Model Number |
DCB00 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/28/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Section a2, a4 and a5: unknown/ not provided.Patient information cannot be provided due to personal data privacy legislation/policy.Section d6a: implant date: not applicable, as lens was removed/replaced in the initial surgery.Section d6b: explant date: not applicable, as lens was removed/replaced in the initial surgery.Section e1: initial reporter first name: information unknown/not provided.Section e1: email address: unknown/not provided.Section e1: telephone number: unknown, information not provided.Section h3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that intraocular lens (iol) was broken when trying to inject into the eye capsule bag.Iol was removed.There was no reported patient injury.No further information available.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d9: device available for evaluation: yes.Date returned to manufacturer: 9 feb 2026.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification was performed on the suspect product and found the lens was coated in viscoelastic residue.The lens was cleaned revealing that the lens was partially cut.A scratch was also observed on the lens.No further evaluation was performed.The complaint issue lens damaged was not identified during product evaluation.The observed cosmetic issues is similar to the reported complaint issue.However, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Manufacturing record evaluation: based on the manufacturing records review, and historical complaint review, there is no indication of a product malfunction or product quality deficiency.The manufacturing records review for this production order (po) shows that the units were released within specification.No nc/er was found as part of this manufacturing records review.For complaint history review, a search of complaints related to this production order (po) was performed in the system.The search revealed that no other complaints were received for this po.No nonconformity report, documentation or labeling changes, and escalations are required.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Search Alerts/Recalls
|