• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AMO PUERTO RICO MFG. INC. TECNIS SIMPLICITY; INTRAOCULAR LENS Back to Search Results
Model Number DCB00
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Section a2, a4 and a5: unknown/ not provided.Patient information cannot be provided due to personal data privacy legislation/policy.Section d6a: implant date: not applicable, as lens was removed/replaced in the initial surgery.Section d6b: explant date: not applicable, as lens was removed/replaced in the initial surgery.Section e1: initial reporter first name: information unknown/not provided.Section e1: email address: unknown/not provided.Section e1: telephone number: unknown, information not provided.Section h3: the device was not returned for evaluation; therefore, a failure analysis of the complaint device cannot be completed.A review of records including device history and complaint trending will be performed.Upon completion of the review, if there is any further relevant information a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that intraocular lens (iol) was broken when trying to inject into the eye capsule bag.Iol was removed.There was no reported patient injury.No further information available.
 
Additional Manufacturer Narrative
Section d9: device available for evaluation: yes.Date returned to manufacturer: 9 feb 2026.Section h3: device evaluated by manufacturer: yes.Device evaluation: visual inspection under magnification was performed on the suspect product and found the lens was coated in viscoelastic residue.The lens was cleaned revealing that the lens was partially cut.A scratch was also observed on the lens.No further evaluation was performed.The complaint issue lens damaged was not identified during product evaluation.The observed cosmetic issues is similar to the reported complaint issue.However, based on the complaint investigation results the complaint issue could not be confirmed to be related to the manufacturing or design process.Manufacturing record evaluation: based on the manufacturing records review, and historical complaint review, there is no indication of a product malfunction or product quality deficiency.The manufacturing records review for this production order (po) shows that the units were released within specification.No nc/er was found as part of this manufacturing records review.For complaint history review, a search of complaints related to this production order (po) was performed in the system.The search revealed that no other complaints were received for this po.No nonconformity report, documentation or labeling changes, and escalations are required.Conclusion: as a result of the investigation, there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SIMPLICITY
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO PUERTO RICO MFG. INC.
road 402 north, anasco ind. pk
anasco PR 00610
Manufacturer Contact
somyata nagpal
31 technology drive
irvine, CA 92618
7142478552
MDR Report Key23795825
Report Number3012236936-2025-000329
Device Sequence Number18761154
Product Code HQL
UDI-Device Identifier05050474636118
UDI-Public(01)05050474636118(17)270920
Combination Product (Y/N)N
Initial Reporter CountryMY
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/31/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberDCB00
Device Catalogue NumberDCB0000215
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/29/2025
Supplement Date Received by Manufacturer03/03/2026
Initial Report FDA Received Date12/12/2025
Supplement Report FDA Received Date03/31/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured09/20/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-