| Model Number |
471171-15 |
| Medical Device Problem Code |
Material Split, Cut or Torn (4008)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/26/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The instrument involved in this complaint has not been received and/or tested by failure analysis as of the date of this report.If the product is received and evaluated, a supplemental mdr will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that the customer experienced an issue with an intuitive product.The customer reported event has no report of patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
Intuitive surgical, inc.(isi) did receive the 8mm micro bipolar forceps instrument to perform failure analysis.The instrument was analyzed and it was found to have a damaged grip cable.Additional findings : the instrument was found to have damage of the conductor wire insulation at the yaw pulley.There were no material missing and the conductor wires were exposed.The instrument passed an electrical continuity test.The probable root cause of cable breakage, whether partial or full, is attributed to damage to the cable through external collisions, during use, or during reprocessing.The probable root cause of damaged conductor wire insulation is attributed to damage from mechanical impact.Accidental drops of the instrument or inadvertent collisions with other instruments or hard surfaces during handling or usage can damage the insulation.
|
| |
|
Event or Problem Description
|
|
Refer to h11 for follow-up information.
|
| |
|