| Catalog Number |
123516A |
| Medical Device Problem Code |
Material Rupture (1546)
|
| Health Effect - Clinical Code |
Exposure to Body Fluids (1745)
|
| Date of Event |
12/08/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that, the water in the foley catheter balloon could not be filled with water normally and appeared to had ruptured.No medical intervention was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event was inconclusive because no sample was returned.It is unknown whether the device had met relevant specifications.It was unknown whether the product had caused the reported failure.The potential root cause for this failure could be user related (example: contact with sharp object)/ exposure to petrolatum based products/ mechanical failure/operator error).A review of the device labelling has been conducted for investigation purposed.The labeling and instructions for use were found to be adequate.A review of the device history record did not show any problems or condition that would have contributed to the reported issue.Complete initial reporter facility name: (b)(6) hospital.This report references a us equivalent device.The us udi equivalent for this product number is used.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|
Event or Problem Description
|
|
It was reported that, the water in the foley catheter balloon could not be filled with water normally and appeared to had ruptured.No medical intervention was reported.Per follow up information received via ibc on 07jan2026, stated that no patient harm was reported.
|
| |
|