The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.Udi format updated with available product information per gudid.H11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
The reported event is confirmed manufacturing related.Received one (1) used all silicone foley catheter without original packaging.Visual inspection noted that there was a hole in the drainage funnel measuring (0.055in).Therefore, the reported event is confirmed and does not meet specifications per which states catheter leaks, valve leaks, balloon perforation, lumen to lumen, prematurely deflated and dislodged balloon are not allowed, and must inflate and deflate with the use of a syringe.The root cause for this failure, per capa 12182448, is the silicone injected into the forming cavity is entering pre cured, which causes overheating in the gate area or injection point, as a result, the breaking point between the gate and the molded part is not properly generated, which prevents automatic separation.This requires manual detachment, increasing the risk of perforation at the adapter injection point due to the high temperature in the gate caused by the lack of flow in the jacket.Risk review, labeling review, and dhr review are not required as event has already being investigated per capa 12182448.Corrections: d, f, h.Udi format updated with available product information per gudid.H11: sections a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|