• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. MAKO ROBOTIC ARM 3.1; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. MAKO ROBOTIC ARM 3.1; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number 219999
Medical Device Problem Codes Application Program Problem: Parameter Calculation Error (1449); Incorrect, Inadequate or Imprecise Result or Readings (1535)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/27/2025
Type of Reportable Event Malfunction
Event or Problem Description
The following information was provided: an issue occurred where distance re-measurement was not captured correctly during cup impaction.After inserting the acetabulum with the trident ii attached to the inline offset impactor, the mako arm was attached using the free arm.At this point, the cup was displayed on the screen significantly more proximally than the reaming area.After releasing the free arm and capturing the distance remaining while pressing the end effector, the remaining distance was displayed as 1mm.Upon impaction, the screen display appeared to align within the reaming area.After several impactions, capturing the distance using the same procedure showed a remaining distance of 5mm.Furthermore, after performing several more impactions from that point, the cup on the screen appeared to slide slightly distally.After the final impaction, capturing the position using the same steps showed a remaining distance of 11mm.After removing the impactor from the cup and the surgeon visually confirmed, the cup was firmly seated against the acetabulum.The surgical result was target 40/20, capture 39/22.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAKO ROBOTIC ARM 3.1
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer (Section G)
MAKO SURGICAL CORP.
3365 enterprise ave
weston FL 33331
Manufacturer Contact
sanjana talathi
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key23797847
Report Number3005985723-2025-00603
Device Sequence Number5546176
Product Code OLO
UDI-Device Identifier07613327395280
UDI-Public07613327395280
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/08/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number219999
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/27/2025
Supplement Date Received by Manufacturer04/10/2026
Initial Report FDA Received Date12/13/2025
Supplement Report FDA Received Date05/08/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/07/2022
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-