Catalog Number ESBF2514C103E
Medical Device Problem Code Malposition of Device (2616)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 11/28/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
An endurant stent graft system was implanted during the endovascular treatment of an abdominal aortic aneurysm.It was reported that approximately 4 years and 11 months post-index procedure, a type ib endoleak was identified involving the right limb (etlw1613c93e).The common iliac artery anatomy had dilated, causing the limb to lose its seal and migrate proximally into the aneurysm sac.An intervention was performed two days later, during which an endurant limb and an endurant cuff were implanted.It was said that although there was no type ia endoleak, it was noted that there was additional space from the top of the esbf2514c103e graft to the renal arteries, as the original implant had landed lower than planned.The physician noted that there was no concerns with the bifurcate stent graft, but wanted to implant the cuff to prevent further issues.According to the physician, the cause of the type ib endoleak was most likely due to a short common iliac artery landing zone, combined with progressive dilation of the common iliac artery, which resulted in loss of seal over time.No additional sequelae were reported, and the patient is fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT IIS BIFURCATED STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TR
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
Manufacturer (Section G)
MEDTRONIC IRELAND
parkmore business park west
galway
EI  
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key23800993
Report Number9612164-2025-06182
Device Sequence Number7596212
Product Code MIH
UDI-Device Identifier00643169439986
UDI-Public(01)00643169439986(17)221112
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/12/2022
Device Catalogue NumberESBF2514C103E
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/09/2025
Initial Report FDA Received Date12/15/2025
Date Device Manufactured11/12/2020
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
Patient SexFemale
-
-