According to the complainant the device will not be returned for investigation.We are unable to determine if any product condition could have contributed to the reported medical event and hyperglycaemia.No lot release records were reviewed, as the product lot number could not be provided.Locked down smartphone: phone_control_ios.Omnipod software app version: 2.0.4.Operating system: 18.7.Hardware: iphone15.3.Cgm sensor type: g6.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.High blood glucose is a common symptom for people with diabetes (glucose monitoring data from people with diabetes indicate that on average, they can experience blood glucose levels above 250 mg/dl for 14-25% of the time[1][2][3].), and it would be challenging to speculate on a cause for the complaints without receiving the devices back for an engineering investigation.[1] beck rw, bergenstal rm, cheng p, kollman c, carlson al, johnson ml, rodbard d.The relationships between time in range, hyperglycemia metrics, and hba1c.J diabetes sci technol 2019; 13:614-626 [1] welsh jb, derdzinski m, parker as, puhr s, jimenez a, walker t.Real-time sharing and following of continuous glucose monitoring data in youth.Diabetes ther 2019; 10:751-755 [1] puhr s, derdzinski m, welsh jb, parker as, walker t, price da.Real-world hypoglycemia avoidance with a continuous glucose monitoring system's predictive low glucose alert.Diabetes technol ther 2019;21:155-158.
|
|
It was reported that the patient had attended the emergency department with hyperglycaemia on (b)(6) 2025.The patient's blood glucose levels reached 700 mg/dl while wearing a pod.The patient reported that their phone had crashed and become non-functional, which included loss of access to the omnipod application.The patient did not have a backup method for insulin delivery.Symptoms reported include pallor, vomiting, weakness and feeling generally unwell.The patient was treated with intravenous insulin.The pod was removed at the hospital and discarded.The patient was released later the same day.
|