• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INSULET CORPORATION OMNIPOD 5 CONTROLLER; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INSULET CORPORATION OMNIPOD 5 CONTROLLER; ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP Back to Search Results
Model Number PT-002343
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Codes Hyperglycemia (1905); Vomiting (2144); Pallor (2468)
Date of Event 11/14/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
According to the complainant the device will not be returned for investigation.We are unable to determine if any product condition could have contributed to the reported medical event and hyperglycaemia.No lot release records were reviewed, as the product lot number could not be provided.Locked down smartphone: phone_control_ios.Omnipod software app version: 2.0.4.Operating system: 18.7.Hardware: iphone15.3.Cgm sensor type: g6.Please note, the device identifiers are captured as reported by the complainant and may not align with the device configuration reported in this section as this data is pulled from our cloud based on the reported date of event.High blood glucose is a common symptom for people with diabetes (glucose monitoring data from people with diabetes indicate that on average, they can experience blood glucose levels above 250 mg/dl for 14-25% of the time[1][2][3].), and it would be challenging to speculate on a cause for the complaints without receiving the devices back for an engineering investigation.[1] beck rw, bergenstal rm, cheng p, kollman c, carlson al, johnson ml, rodbard d.The relationships between time in range, hyperglycemia metrics, and hba1c.J diabetes sci technol 2019; 13:614-626 [1] welsh jb, derdzinski m, parker as, puhr s, jimenez a, walker t.Real-time sharing and following of continuous glucose monitoring data in youth.Diabetes ther 2019; 10:751-755 [1] puhr s, derdzinski m, welsh jb, parker as, walker t, price da.Real-world hypoglycemia avoidance with a continuous glucose monitoring system's predictive low glucose alert.Diabetes technol ther 2019;21:155-158.
 
Event or Problem Description
It was reported that the patient had attended the emergency department with hyperglycaemia on (b)(6) 2025.The patient's blood glucose levels reached 700 mg/dl while wearing a pod.The patient reported that their phone had crashed and become non-functional, which included loss of access to the omnipod application.The patient did not have a backup method for insulin delivery.Symptoms reported include pallor, vomiting, weakness and feeling generally unwell.The patient was treated with intravenous insulin.The pod was removed at the hospital and discarded.The patient was released later the same day.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNIPOD 5 CONTROLLER
Common Device Name
ALTERNATE CONTROLLER ENABLED INSULIN INFUSION PUMP
Manufacturer (Section D)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer (Section G)
INSULET CORPORATION
100 nagog park
acton MA 01720
Manufacturer Contact
derek heyl
100 nagog park
acton, MA 01720
9786007000
MDR Report Key23801079
Report Number3004464228-2025-62251
Device Sequence Number9359940
Product Code QFG
UDI-Device Identifier10385083000459
UDI-Public(01)10385083000459
Combination Product (Y/N)N
PMA/510(K) Number
K231826
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 12/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Model NumberPT-002343
Device Catalogue NumberM009-2S-AP
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/15/2025
Initial Report FDA Received Date12/15/2025
Is the Device Labeled for Single Use? (Y/N) No
Usage of Device Initial
Patient Sequence Number1
Patient Age24 YR
Patient SexFemale
-
-