Additional information was added in h.3.,h.6.And h.11.Service history was reviewed for the system.There was no service record (relevant to the reported event), found.There was no previous service prior to the reported event (other than installation), per the service record (sr).All surgical systems are verified to meet specifications after installation and customer initiated servicing, in accordance with the applicable procedures.Based on the assessment, the product met release criteria.Based on the information obtained, the root cause of the reported event remains inconclusive.A review for ¿complaints reported against this serial number¿ was performed.A relevant complaint was found and reviewed as part of this investigation.This complaint was determined to be relevant, due to the same time frame.The reported event revealed an issue with the scanner shot positioning, resulting in the console being replaced.The reported ¿iris transillumination¿ is categorized as an adverse event tied to ¿iris defect.¿ the issue is seen when viewed with a slit lamp (post-operatively) and shows where pigment has been lost.According to clinical information, ¿these defects are generally painless, and patients are typically asymptomatic.¿ there was no service record for the system (relevant to the reported event), found.Therefore, based on the information obtained, the root cause of the reported event remains inconclusive.The manufacturer internal reference number is: (b)(4).
|