| Catalog Number |
VCP496ZH |
| Medical Device Problem Codes |
Break (1069); Material Separation (1562)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
01/01/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Date sent to the fda: 12/15/2025.H6 component code: g07002 - pending evaluation of returned device.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.Attempts are being made to obtain the following information.To date no response has been provided.If further details are received at a later date a supplemental medwatch will be sent: when were the needles detached/pulled off from the sutures (in the package, during removal from package, during handling prior to use on patient or during use on the patient)? please specify.Please provide the source or name of person providing answers to follow-up questions: a device has been received, however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a skin closure procedure on an unknown date and suture was used.It was reported by the sales rep that during a robotic skin closure, the needles are not attached to the suture.No further information provided.There were no patient consequences reported.Additional information was requested.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h3, h6.Investigation summary: the product was returned to ethicon for evaluation.One representative unopened sample was received for analysis.Product code vcp496zh.Upon initial inspection of the samples, no external damages were observed on the external packet.In order to evaluate the conditions of the returned samples, the packet was opened, and no defects were detected.The swage and attachment area were noted to be as expected.The needle was intact and there was no damage, or breakage on the body, tip, or swage area was observed during the evaluation.The suture was dispensed without problems and examined along the strand to detect any issue related and no defects were observed during evaluation.The sample was tested by resistance test to the needle and met the requirements.A functional test was performed using instron equipment, and the pull force results were above the minimum requirements.The event described could not be confirmed as the device performed without any defect noted.It should be noted that as part of our quality process, each batch is randomly inspected and functionally tested during manufacturing to ensure the device meets the required specifications prior to shipment.The event described could not be confirmed as no product defect was identified, there may have been other circumstances or issues that occurred during the use of the device that could not be replicated during the laboratory analysis.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post-market surveillance.
|
| |
|
Search Alerts/Recalls
|
|