SMITH & NEPHEW ORTHOPAEDICS LTD FEMORAL HEAD 48MM; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/METAL, RESURFACING
|
Back to Search Results |
|
| Catalog Number |
74123148 |
| Medical Device Problem Code |
Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Foreign Body Reaction (1868); Pain (1994); Osteolysis (2377); Joint Laxity (4526); Metal Related Pathology (4530)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that, after a thr surgery performed in 2015, the patient has experienced hip pain and internal issues with surrounding area from apparent metallosis.It is unknown if any treatment has been provided or if a revision surgery has been scheduled.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: internal complaint reference: (b)(4).H3, h6: this complaint was opened by smith+nephew to document a patient complication related to a product problem with a smith+nephew device.The reported issue(s) relate to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
|
| |
|
Additional Manufacturer Narrative
|
|
H11: additional information in a1, a2, a3, a4, b5, b6 and d10.
|
| |
|
Event or Problem Description
|
|
It was reported that, after a thr surgery performed on (b)(6) 2015, the patient has experienced hip pain, instability, with surrounding area from apparent metallosis.No revision surgery and no additional treatment was indicated by the hcp.
|
| |
|
Additional Manufacturer Narrative
|
|
A1-a4, b7, h3, h6.As of today, the devices, all of which were used in treatment, remain implanted and therefore cannot be tested.A review of the historical complaints data for the alleged devices was performed using batch numbers, part numbers and the reported failure modes to evaluate patterns of repeated failures or defects.Similar complaints have been identified for the cup and head.This failure will continue to be monitored via routine trending.In the absence of the actual devices, the production records were reviewed for the devices reportedly involved in this incident.All released devices involved met manufacturing specifications upon release for distribution.The review of the product ifu found adequate warnings and precautions in relation to the alleged failure modes.A risk management review was performed.The alleged failure modes and associated risks have been anticipated within the risk file and the anticipated risk level is still adequate.No further actions are required at this time.A review of historic escalation actions related to the products and similar complaint events was performed.Following the review, no prior applicable escalation actions were identified.No further escalation actions are required.The available medical documents were reviewed.The osteolysis, hypointense synovial thickening and trace joint effusion are consistent with an adverse local tissue reaction/particle disease.However, based on the limited information provided the reported metallosis cannot be confirmed.The patient impact is determined to be the reported pain, instability, osteolysis, hypointense synovial thickening and trace joint effusion.A definitive root cause cannot be determined.Specific factors known to contribute to the alleged fault are excessive physical activity levels, unreasonable stress on replacement system, excessive patient weight, trauma to the joint replacement, loosening of components may increase production of wear particles and accelerate damage to the bone.Should the devices or additional information be received, the complaint will be reopened.Based on this investigation, the need for corrective and preventative actions is not indicated.
|
| |
|
Search Alerts/Recalls
|
|
|