Model Number HEM3335
Medical Device Problem Code Mechanics Altered (2984)
Health Effect - Clinical Codes Hemorrhage/Blood Loss/Bleeding (1888); Tissue Breakdown (2681)
Date of Event 05/28/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter during a procedure for the treatment of hemorrhoids, the device failed to properly place staples, resulting in a significant section of the rectal mucosa being cut.The issue caused substantial rectal bleeding, specifically in about 2/3 to 3/4 of the circumference of the mucosal area.The surgeon attempted to control the bleeding by suturing the mucosal edges and performing targeted hemostasis over the bleeding section.
 
Additional Manufacturer Narrative
Correction: b5 medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Event or Problem Description
According to the reporter, during a procedure for the treatment of hemorrhoids, the device failed to place staples, performed only the section, resulting in a significant section of the rectal mucosa being cut.The issue caused substantial rectal bleeding, specifically in about 2/3 to 3/4 of the circumference of the mucosal area.The surgeon attempted to control the bleeding by suturing the mucosal edges and performing targeted hemostasis over the bleeding section.
 
Additional Manufacturer Narrative
Additional information: g3, h6 h3 evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted that the visual inspection of the staple guide noted the instrument was fully applied.Further inspection noted that the green bar was not visible in the indicator window and the safety feature was not engaged.The staple guide was observed to be attached to the instrument with no damage to the staple guide.After actuating the handle, the pusher fingers appeared to be intact and inspection under the microscope displayed no damage to the knife blade.{anvil}.The anvil of the instrument was observed to be separated from the instrument.Further inspection of the anvil cutting ring showed deep traces of knife impression and the ring was received completely severed.An acceptable gap between anvil and staple guide was observed when fully approximated.It was reported that staples did not deploy.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EEA
Common Device Name
STAPLE, IMPLANTABLE
Manufacturer (Section D)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer (Section G)
COVIDIEN SURGICAL
60 middletown ave
north haven CT 06473
Manufacturer Contact
mike bocchino
8200 coral sea st ne
mounds view, MN 55112-4391
8006464633
MDR Report Key23803967
Report Number1219930-2025-05774
Device Sequence Number15777238
Product Code GDW
UDI-Device Identifier10884521047518
UDI-Public(01)10884521047518(17)281130(10)N3M1528Y(20)01
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K083781
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional,Company Representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/13/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberHEM3335
Device Catalogue NumberHEM3335
Device Lot NumberN3M1528Y
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/05/2025
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/20/2025
Supplement Date Received by Manufacturer01/12/2026
02/04/2026
Initial Report FDA Received Date12/15/2025
Supplement Report FDA Received Date01/12/2026
02/13/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured12/01/2023
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age47 YR
Patient SexMale
-
-