• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR W/OCORD; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX ADVANCE BR W/OCORD; FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 222297
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated, and no conclusion can be drawn at this time.
 
Event or Problem Description
It was reported that during a rotator cuff repair surgery, the 5.5 healix advance br w/ocord anchor device was broken off.It was reported that all broken parts were removed.Another device was used to complete the surgery.Additionally, it was reported that there was foreign matter observed on the vapr premiere 90 electrode device.Another device was used to complete the surgery.There were no adverse consequences to the patient.No additional information could be provided.Report 2 of 2 for (b)(4).
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the product was returned to depuy synthes for evaluation.Visual inspection found that the device had the anchor still attached to the inserter, the anchor was cracked and broken vertically.No other anomalies were noted.The overall complaint was confirmed as the observed condition of 5.5 healix advance br w/ocord would have contributed to the complained device issue.Based on the investigation findings, the potential cause is traced to the procedural variables, such handling of the device or product interaction during procedure.The possible root cause can be associated with off axis insertion and levering during insertion.As per the instructions for use, axial misalignment or levering with the anchor upon insertion may result in anchor fracture.It has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history review: there was no non-conformance regarding this lot.
 
Additional Manufacturer Narrative
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.H11 additional narrative: the actual device was returned for evaluation.However, since the investigation is still on-going, the assignable root cause could not be determined at this time.Once the investigation has been completed, a follow-up report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE BR W/OCORD
Common Device Name
FASTENER, FIXATION, BIODEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
elba bello
325 paramount drive
raynham, MA 02767
9083863534
MDR Report Key23806067
Report Number1221934-2025-04772
Device Sequence Number9389958
Product Code MAI
UDI-Device Identifier10886705021338
UDI-Public(01)10886705021338(10)10249L(17)270630
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K120078
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number222297
Device Lot Number10249L
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2025
Supplement Date Received by Manufacturer01/05/2026
01/06/2026
Initial Report FDA Received Date12/15/2025
Supplement Report FDA Received Date01/06/2026
01/07/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured07/10/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-