| Model Number |
PM1172 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/02/2025
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that the device was explanted and replaced as it is subject to the assurity, endurity inhomogeneous epoxy mixing field safety correction action which applies to a subset of devices distributed and implanted outside of the united states.The patient was stable and there were no adverse consequences.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned due to advisory with no allegation of a malfunction.Visual inspection of the device and header attachment area did not find any anomalies.Interrogation of the device indicated battery voltage and current within normal range when received.Based on this information, the device was found to communicate appropriately with a merlin programmer and has not reached the elective replacement indicator (eri).
|
| |
|