• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: IRIDEX CORPORATION G-PROBE RFID

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

IRIDEX CORPORATION G-PROBE RFID Back to Search Results
Lot Number 900910
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Eye Injury (1845)
Date of Event 02/21/2024
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported to iridex that while using the g-probe rfid, the patient experienced 2 small burn spots at limbus post procedure and was given lido gel for treatment.No other injuries reported.Treatment and power settings not provided.The g-probe was discarded the probe and continued treatment with a different g-probe without any failures detected.The g-probe was was not returned to iridex for further investigation.Conjunctival burns are a known potential harm associated with the use of laser delivery in ophthalmic indications.It is listed in the ifu included in mp3 delivery device.No information regarding the patient, patient recovery, treatment parameters used were provided to iridex to have a conclusive finding.In accordance with standard protocol, iridex will proceed with processing the complaint based on the information currently available.Another factor to be taken into account for the patient injury reported is that occasionally, the tip of the probe will become contaminated with blood or tissue, and the contaminants on the tip can become hated when exposed to laser energy.Typically, conjunctival burns resolve within 24 to 48 hours.Therefore, unless either medical intervention or permanent impairment are reported, scleral burns are not deemed a serious adverse event, because conjunctival burns typically: do not result in life-threatening illness or injury, do not result in permanent impairment of a body structure or a body function, do not require hospitalization or prolongation of patient hospitalization, do not require medical or surgical intervention to prevent life-threatening illness or injury or permanent impairment to a body structure or a body function.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G-PROBE RFID
Common Device Name
G-PROBE RFID
Manufacturer (Section D)
IRIDEX CORPORATION
1212 terra bella avenue
mountain view CA 94043 1824
Manufacturer (Section G)
IRIDEX CORPORATION
1212 terra bella avenue
mountain view CA 94043 1824
Manufacturer Contact
bill hyatt
1212 terra bella avenue
mountain view, CA 94043-1824
6506058727
MDR Report Key23809700
Report Number2939653-2024-00025
Device Sequence Number7707678
Product Code GEX
UDI-Device Identifier00813125015237
UDI-Public00813125015237
Combination Product (Y/N)N
Initial Reporter StateCT
Initial Reporter CountryUS
PMA/510(K) Number
K143154
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Lot Number900910
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 02/21/2024
Initial Report FDA Received Date12/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-