MAUDE Adverse Event Report: MERIT MEDICAL SYSTEMS, INC. PRELUDE®; VASCULAR DILATOR, SINGLE-USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MERIT MEDICAL SYSTEMS, INC. PRELUDE®; VASCULAR DILATOR, SINGLE-USE Back to Search Results
Catalog Number PSI-6F-23-038/D
Medical Device Problem Codes Fracture (1260); Material Separation (1562)
Health Effect - Clinical Code Hemorrhage/Blood Loss/Bleeding (1888)
Date of Event 11/23/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The suspect device is expected to return for evaluation.A follow up will be submitted when the evaluation is complete.
 
Event or Problem Description
The account alleges that during nursing care, the patient was turned on their side and a large pool of blood was observed in the bed.The catheter was examined and the side arm had broken off.No patient injury to report.
 
Additional Manufacturer Narrative
The suspect device was returned for evaluation.The complaint could not be confirmed.The root cause could not be determined.A review of the device history and complaint database could not be performed since the lot number was not provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRELUDE®
Common Device Name
VASCULAR DILATOR, SINGLE-USE
Manufacturer (Section D)
MERIT MEDICAL SYSTEMS, INC.
1600 merit parkway
south jordan UT 84095
Manufacturer (Section G)
MERIT MEDICAL SYSTEMS, INC.
1600 merit parkway
south jordan UT 84095
Manufacturer Contact
adrian cole
1600 merit parkway
south jordan, UT 84095
8012084225
MDR Report Key23810432
Report Number1721504-2025-00729
Device Sequence Number12689604
Product Code DRE
UDI-Device Identifier00884450614617
UDI-Public00884450614617
Combination Product (Y/N)Y
Initial Reporter CountrySZ
PMA/510(K) Number
K070159
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/15/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Catalogue NumberPSI-6F-23-038/D
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2025
Supplement Date Received by Manufacturer01/14/2026
Initial Report FDA Received Date12/15/2025
Supplement Report FDA Received Date01/15/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Is the Device Labeled for Single Use? (Y/N) Yes
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-