Model Number 1020000
Medical Device Problem Codes Thermal Decomposition of Device (1071); Melted (1385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/17/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received a work order reporting that the everflo device has issues with power cord and integrated canister.During the evaluation of the device at the third-party service center, the device was visually inspected and found the power cord melted at one end and sieve columns were failing.The device power cord and sieve columns were replaced.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERFLO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
7247330200
MDR Report Key23810621
Report Number2518422-2025-112505
Device Sequence Number15792687
Product Code CAW
UDI-Device Identifier00606959014954
UDI-Public(01)00606959014954
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K061261
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Distributor
Initial Reporter Occupation 505
Type of Report Initial
Report Date (Section B) 12/15/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model Number1020000
Device Catalogue Number1020000
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/17/2025
Initial Report FDA Received Date12/15/2025
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured03/06/2024
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-