| Model Number |
CDS0706-XT |
| Medical Device Problem Codes |
Difficult or Delayed Positioning (1157); Incomplete Coaptation (2507)
|
| Health Effect - Clinical Code |
Mitral Valve Insufficiency/ Regurgitation (4451)
|
| Date of Event |
11/19/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
|
| |
|
Event or Problem Description
|
|
It was discussed of needed, the transeptal would need to be potentially positioned in the amplatzer device, after several attempts a transeptal puncture was achieved posterior to the amplatzer device, and a measurement of 4.3cm above the valve plane was noted.Upon insertion of the sgc and cds, and steering down to the valve, in order to achieve perpendicularity, gaining height was needed.(p) knob was applied and adjustment of (m) knob was done to gain adequate height to position clip above the level of the leaflets.Upon crossing the valve after correct arm orientation was achieved via 3d enface view and lvot view, a grasp was made and successfully deployed, after thorough leaflet assessment, on the medial aspect of the a2/p2 portion of the valve, with some reduction in mr grade to 3+ down from 4+ initially.A mean gradient of 2mmhg was noted, the v wave was lowered to 40mmhg,.There was still mr lateral to the clip and the team discussed the risks/benefits of adding a 2nd clip lateral to the 1st.A 2nd xtw was directed and the same maneuvers were done to successfully grasp the lateral portion of the leaflets, after a thorough leaflet assessment.Upon deployment of the clip due to height issues the clip was deployed per the ifu and the clip was free from the delivery system.However, the shaft of the delivery system was still in contact with the successfully deployed clip.A unstraddle of the cds was done to attempt to create space from the clip and a slight removal of (m) knob was also done, at that point the cds shaft was interacting with the clip and then it was observed under flouro and echo that there was an slda of the anterior leaflet.A 3rd clip was steered down to attempt to stabilizer the slda clip #2.After successful attempt to grasp the leaflets with the 3rd clip and a thorough leaflet assessment was done, it was deployed per the ifu, and the cds was removed out of the la.After that was pulled into the sgc, and access was maintained in the la, it was noted that the 3rd clip was now moving in tandem with the slda clip and was indeed an additional slda off the anterior leaflet.The 1st clip remained intact and had adequate leaflets in both arms.4+ mr was present with a gradient of 3mmhg, and v wave was 50mmhg, better than baseline.It was decided to abort the procedure and see how the patient was after anesthesia.Hemostasis was achieved in the groin and pt was moved to cardiac icu for further evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record revealed no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar complaints reported from this lot.Based on available information, the reported difficult positioning was due to patient condition.The cause of the reported slda was unable to be determined.The reported unchanged mr seems to be related to the reported slda.The reported patient effect of mitral regurgitation, as listed in the mitraclip system instructions for use, is a known possible complication associated with mitraclip procedures.The reported serious injury/ illness/ impairment was a result of case-specific circumstances.There is no indication of a product quality issue with respect to manufacture, design, or labeling.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2025, a mitraclip procedure was performed to treat functional mitral regurgitation (mr) grade 4+.The patient had a pre-existing non-abbott left atrial appendage occlusion device and a pre-existing amplatzer septal occlusion device.After several attempts, a transeptal puncture was achieved posterior to the amplatzer, with the puncture height measured at 4.3 cm above the valve plane.Following this, the sgc and cds were introduced, and additional height was required to achieve perpendicularity to the mitral valve; this was accomplished using the posterior (p) knob and adjustments to the medial (m) knob to adequately position the clip above the leaflet level.Once proper arm orientation was confirmed using 3d en face and lvot views, a first clip was successfully grasped and deployed on the medial a2/p2 segment, reducing the mitral regurgitation from 4+ to 3+ with a mean gradient of 2 mmhg and reduction of the v-wave to 40 mmhg.Because residual lateral mr persisted, the team proceeded with a second xtw clip positioned lateral to the first.Although leaflet grasping and assessment were satisfactory, the second clip encountered deployment difficulties related to insufficient height, and while the clip released appropriately per the ifu, the cds shaft remained in contact with the newly deployed clip.Efforts to unstraddle the cds and adjust the m knob were made; however, interaction between the cds shaft and the device resulted in a single-leaflet device attachment (slda) of the anterior leaflet, confirmed by fluoroscopy and echocardiography.A third clip was then advanced in an effort to stabilize the slda clip.After successful leaflet grasping and thorough assessment, the third clip was deployed per the ifu, and the cds was withdrawn into the sgc while maintaining left atrial access.Shortly afterward, it became apparent that the third clip was now moving in tandem with the slda clip, indicating a second slda involving the anterior leaflet, while the first clip remained well-positioned with durable leaflet insertion.At this stage, the patient exhibited recurrent 4+ mr with a mean gradient of 3 mmhg and a v-wave of 50 mmhg, which, while improved from baseline, represented a significant residual lesion.Given multiple device-related complications¿specifically slda of both the second and third clips¿and persistent severe mr, the team elected to abort the procedure and assess the patient following anesthesia recovery.Hemostasis was obtained at the groin access site, and the patient was transferred to the cardiac icu for ongoing evaluation.
|
| |
|