| Model Number |
TFL-FBX150BS |
| Medical Device Problem Code |
Flare or Flash (2942)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
11/17/2025
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
It was reported the ball tip single use fiber exhibited a laser beam was observed emitting from the mid-section of the fiber.The issue occurred during a therapeutic retrograde intrarenal surgery.There were no reports of patient harm.
|
| |
|
Additional Manufacturer Narrative
|
|
This supplemental report is being submitted to provide the results of the final investigation.Updated fields: d8, d9, h4, h6.The device was returned to olympus for inspection and the customer allegation was confirmed due to component failure in the fiber coating material.A root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
|
| |
|
Event or Problem Description
|
|
No new additional information provided by the customer.
|
| |
|