• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC NEUROMODULATION CORPORATION PRECISION?; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC NEUROMODULATION CORPORATION PRECISION?; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number SC-9218-15
Medical Device Problem Code High impedance (1291)
Health Effect - Clinical Code Pain (1994)
Date of Event 05/15/2025
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that the patient experienced inadequate spinal cord stimulation (scs) as a result of high impedances displayed on the lead.Device programming optimization was attempted; however, the issue persisted.The patient underwent a revision procedure, where troubleshooting found there was an unstable connection between the lead adapter and the non-boston scientific lead.Therefore, the lead adapter and non-boston scientific lead were explanted, and a new boston scientific lead was implanted.Postoperatively, the patient was content and received full stimulation coverage.The device will not be returned as it was retained by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION?
Common Device Name
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION CORPORATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
road 698, lot no. 12
dorado PR 00646 -260
*   00646-2602
Manufacturer Contact
erik sherburne
25155 rye canyon loop
valencia, CA 91355
7632920920
MDR Report Key23813490
Report Number3006630150-2025-11456
Device Sequence Number9409223
Product Code LGW
UDI-Device Identifier08714729888734
UDI-Public(01)08714729888734(17)241004(10)7074847
Combination Product (Y/N)N
Initial Reporter CountryGM
PMA/510(K) Number
P030017
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2023
Device Explanted Year2025
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 12/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date10/04/2024
Device Model NumberSC-9218-15
Device Catalogue NumberSC-9218-15
Device Lot Number7074847
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/04/2025
Initial Report FDA Received Date12/16/2025
Date Device Manufactured10/05/2022
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient SexUnknown
-
-