ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER CENTURION SILVER SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE; UNIT, PHACOFRAGMENTATION Back to Search Results
Catalog Number 8065750469
Medical Device Problem Code Suction Failure (4039)
Health Effect - Clinical Code Capsular Bag Tear (2639)
Date of Event 10/01/2025
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Event or Problem Description
A customer reported that an ophthalmic handpiece exhibited no aspiration in cortex with aspiration and irrigation (i/a) mode.A sudden chamber fluctuation led to a capsular rupture.Emergency surgery was required, followed by secondary implantation.The ia handpiece and tip were replaced.There was no reports of procedure type.The current outcome of the patient was unknown.
 
Additional Manufacturer Narrative
Additional information is provided in section h.6 and h.11.The product under investigation is not a serviceable device.Therefore, a service record review was not performed.There was no product returned for this investigation.Based on the information obtained, the root cause of the reported event is inconclusive.A non-conformance based review of the serial number was performed and did not reveal any potential contributing factors to the reported complaint.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A review for ¿complaints reported against this serial number¿ was performed.No similar complaints were reported for the product serial under investigation.There was no product returned, and additional information was not provided.Therefore, based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.Similar incident/event data was collected for the associated reported events.Quality assurance will continue to perform periodic monitoring for evidence of adverse trending and take further action as appropriate.The manufacturer internal reference number is: (b)(4).H.10 reflects all related report numbers associated with this product event that have been submitted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTURION SILVER SYSTEM, ACCESSORY, OZIL TORSIONAL HANDPIECE
Common Device Name
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key23814929
Report Number2028159-2025-01872
Device Sequence Number10006862
Product Code HQC
UDI-Device Identifier00380657504695
UDI-Public(01)00380657504695(17)340617(11)240619(10)16WCXR
Combination Product (Y/N)N
Initial Reporter CountryMB
PMA/510(K) Number
K121555
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 03/27/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number8065750469
Device Lot Number16WCXR
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/24/2025
Supplement Date Received by Manufacturer03/03/2026
Initial Report FDA Received Date12/16/2025
Supplement Report FDA Received Date03/30/2026
Date Device Manufactured06/19/2024
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient SexUnknown
-
-