BLUE BELT TECHNOLOGIES REAL INTELLIGENCE ROBOTIC DRILL GUARD; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Catalog Number ROB10016
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/25/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).This complaint was opened by smith+nephew to document a product problem associated with a smith+nephew device.The reported problem relates to known risks with the device itself or with its use that are appropriately documented in our risk files.Smith+nephew will continue to monitor trends in accordance with our post-market surveillance process and take necessary action as required if anticipated severity and/or occurrence rates are exceeded.Smith+nephew has no reason to suspect that the product failed to meet any specifications at the time of manufacture.Based on our review of all currently available information, we are unable to identify a definitive root cause.However, as the use of our product cannot be excluded as a potential cause or contributory factor to the reported issue, we are submitting this report in accordance with applicable regulations.If additional investigative findings or information becomes available that alters the conclusions of this report, a follow-up report will be submitted as required.
 
Event or Problem Description
It was reported that, during a cori assisted tka surgery, three (3) real intelligence robotic drill guards were tried and none of them would allow insertion into one (1) real intelligence point probe.The problem was with the reported guards as the tips were scarred.It is unknown if there was any surgical delay.The procedure was resumed with a change in surgical technique (manual procedure).No further complications were reported.
 
Additional Manufacturer Narrative
H3, h6: the real intelligence robotic drill guard, part number rob10016, lot number 20eumc002, intended for treatment was not returned for evaluation.An image was provided.The reported problem was confirmed with a visual inspection.The tip of the guard is scarred.A complaint history review was performed by part number and similar complaints were identified.A review by lot number was performed and no similar complaints were identified.A review of manufacturing records indicates the device met all specifications upon release into distribution.The failure mode and associated risk have been anticipated within the risk file, and the documented risk level is still adequate.A historical review concluded that no prior escalation actions are applicable to the scope of the reported complaint.We have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The most likely cause of this event is associated with wear and tear.Based on the investigation, no containment or corrective action is recommended or required at this time.Should any additional information be received, the complaint will be reopened.The failure mode will continue to be closely monitored through complaint investigation and trended through post market surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAL INTELLIGENCE ROBOTIC DRILL GUARD
Common Device Name
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer (Section G)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
Manufacturer Contact
holly topping
1450 brooks rd.
memphis, TN 38116
5123913905
MDR Report Key23816360
Report Number3010266064-2025-00273
Device Sequence Number10006910
Product Code OLO
UDI-Device Identifier00885556757352
UDI-Public885556757352
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
K193120
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,Health Professional,Company Representative
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/07/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberROB10016
Device Lot Number20EUMC002
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/25/2025
Supplement Date Received by Manufacturer12/31/2025
Initial Report FDA Received Date12/16/2025
Supplement Report FDA Received Date01/07/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured06/18/2020
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Unknown
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PN: ROB10013 / SN: UNKNOWN; PN: ROB10016 / LOT: 1128358; PN: ROB10017 / LOT: 20KCT0007; PN: ROB10024 / SN: (B)(6)