|
Additional information: d9, g3, h3, h6 the complaint allegation was confirmed.One unpackaged ar-1588tb-1 batch 15299791 was received for evaluation.The button was received with a piece of suture, presumably part of the graft.Visual evaluation of the button found that the device was broken in half.Evaluation of both pieces revealed no sharp edges; however, some scratches, small marks, and indentations were most likely caused by other instruments.The evaluation of the returned suture noted fraying at the tips with broken filaments.The condition of the suture tips indicates that the suture was broken, as the ends do not exhibit a clean, uniform cut.Functional tests were not performed due to the device damage.The most likely cause of the reported failure was a patient-specific factor, such as the patient¿s inability or unwillingness to adequately restrict activities or adhere to postoperative instructions during the prescribed healing period.Directions for use dfu-0147-eo revision 4 tightrope® devices c.Contraindications 6.Conditions that tend to limit the patient's ability or willingness to restrict activities or follow directions during the healing period.E.Warnings 9.Postoperatively and until healing is complete, fixation provided by this device should be considered as temporary and may not withstand weight bearing or other unsupported stress.The fixation provided by this device should be protected.The postoperative regimen prescribed by the physician should be strictly followed to avoid adverse stresses applied to the device and bone.
|