On (b)(6) 2025, it was reported by a sales representative via (b)(4) that an ar-8400cds dissector had small pieces of metal introduced into the joint from the shaver.Upon examination of the shaver, there were lines on the actuating portion of the device.This was discovered during a knee arthroscopy.Another device was used to complete the case with no patient harm.
|
|
Additional information: g3, h3, h6 -complaint allegation is not confirmed.Based on the information provided which may include the device (if available and returned), pictures, videos, event description, and any additional information from the field, arthrex was able to conclude a most likely cause.Per dfu-0215-eo- f precautions 6.Do not allow the rotating portion of any device to touch any metallic object, such as a cannula or arthroscope, during use.Damage to both devices is likely.Damage to the device can range from a slight distortion or dulling of the blade/burr edge to actual fracture of the tip in vivo.If such contact does occur, inspect the tip.If there are cracks, fractures or dulling, or if there is any other reason to suspect a blade is damaged, replace it immediately.7.Excessive "side-loading" on the device during use does not improve cutting performance and in extreme cases will cause irreversible damage to the device.The most likely cause is attributed to misuse of the device, including exposure to excessive mechanical forces, misaligned insertion, side loading during use and/or allowing the rotating portion to touch any metallic object during use.
|