• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. SFXSYM PDS+ UNI VIO 24IN 1 SA CT; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. SFXSYM PDS+ UNI VIO 24IN 1 SA CT; SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE Back to Search Results
Catalog Number SXPP1A444
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/19/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that a patient underwent an unknown procedure on (b)(6) 2025 and suture was used.It was reported that the suture was broken when the surgeon attempted to sew the tissue.Changed another one to complete the surgery.There were no patient consequences reported.Additional information was requested.
 
Additional Manufacturer Narrative
Product complaint # (b)(4) date sent to the fda: 12/16/2025 h6 component code: g07002 - pending evaluation of returned device.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.A manufacturing record evaluation was performed for the finished device lot, and no non-conformances were identified.A device has been received; however it has not yet been evaluated.Any further information derived from the evaluation will be submitted in a supplemental 3500a form.
 
Additional Manufacturer Narrative
Product complaint #: (b)(4).Date sent to the fda: 1/13/2026.This report is being submitted pursuant to the provisions of 21 cfr, part 803, part 4 subpart b.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Additional information: h3, h6.H3 investigational summary: the product was returned to eth for evaluation.Visual inspection revelated that one labeled winding former with a used needle suture piece and one empty foil packet were received for analysis.Product code sxpp1a444.During visual inspection of the returned sample, marks that appear to be caused by a surgical instrument were observed on the body needle.The suture was examined, and the end of the suture was noted to be split and cut probably caused by a surgical instrument.In addition, body fluids and crimping marks were found on the strand.The other section of the suture was not returned for analysis.As the suture is absorbable and the duration of exposure of the suture in the environment could not be determined, the functional test could not be performed.As with any device, care should be taken to avoid damage to the strand when handling.Avoid the crushing or crimping action of surgical instruments, such as needle holders and forceps.Please reference the instruction for use for more information.As part of eth quality process all devices are manufactured, inspected, and released to approved specifications.Additional complaint information monitoring for potential safety signals is conducted through complaint trending as part of post market surveillance.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SFXSYM PDS+ UNI VIO 24IN 1 SA CT
Common Device Name
SUTURE, SURGICAL, ABSORBABLE, POLYDIOXANONE
Manufacturer (Section D)
ETHICON INC.
1000 route 202
raritan NJ 08869
Manufacturer (Section G)
ETHICON INC.-JUAREZ
calle durango #2751
ciudad juarez 32575
MX   32575
Manufacturer Contact
orla o'mahony
1000 route 202
raritan, NJ 08869
3035526892
MDR Report Key23819376
Report Number2210968-2025-14128
Device Sequence Number13287999
Product Code NEW
UDI-Device Identifier10705031245235
UDI-Public(01)10705031245235(17)261231(10)105SMJ
Combination Product (Y/N)Y
Initial Reporter CountryCH
PMA/510(K) Number
K141776
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Foreign,Health Professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 12/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberSXPP1A444
Device Lot Number105SMJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2025
Is the Reporter a Health Professional? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 11/19/2025
Supplement Date Received by Manufacturer12/17/2025
Initial Report FDA Received Date12/16/2025
Supplement Report FDA Received Date01/13/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured01/17/2025
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-