• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O. HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O. HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30° Back to Search Results
Model Number WA22306D
Medical Device Problem Code Melted (1385)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/21/2025
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is ongoing.A supplemental report will be submitted when the investigation is completed or if additional information becomes available.
 
Event or Problem Description
It was reported the hf-resection electrode had the loop at the tip melted and cut off.This malfunction occurred during a therapeutic transurethral resection of the prostate procedure.There were no reports of patient harm/involvement.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide the results of the final investigation.The device was returned to olympus for inspection and the reported failure was confirmed.A definitive root cause could not be identified.Based on the results of the investigation, it is likely the following led to the malfunction: the most probable cause for the malfunction was traced to component failure.Olympus will continue to monitor field performance for this device.
 
Event or Problem Description
No additional information provided by the customer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HF-RESECTION ELECTRODE, LOOP, 24 FR., 0.2 WIRE, MEDIUM, 30°
Common Device Name
HF-RESECTION ELECTRODE
Manufacturer (Section D)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ  75124
Manufacturer (Section G)
OLYMPUS MEDICAL PRODUCTS CZECH, SPOL. S R.O.
telickova 457/29
prerov 75124
EZ   75124
Manufacturer Contact
todd brill
801 west park drive
westborough, MA 01581
5082077661
MDR Report Key23820552
Report Number3007960282-2025-00233
Device Sequence Number9963795
Product Code FAS
UDI-Device Identifier14042761051689
UDI-Public14042761051689
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K100275 / K
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,User Facility
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 01/16/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberWA22306D
Device Lot Number1000158012
Was Device Available for Evaluation? Device Returned to Manufacturer
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 11/27/2025
Supplement Date Received by Manufacturer12/25/2025
Initial Report FDA Received Date12/16/2025
Supplement Report FDA Received Date01/16/2026
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured08/29/2024
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-