MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II COLONOVIDEOSCOPE Back to Search Results
Model Number CF-Q180AL
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/18/2025
Type of Reportable Event Malfunction
Event or Problem Description
It was observed that during the device evaluation, the colonovideoscope's image was not displayed.There was no patient involvement.
 
Additional Manufacturer Narrative
The device was returned to olympus for inspection.Based on the results of the investigation, it is likely that corrosion on the connector caused the image to not display.However, a definitive root cause could not be identified.Should additional relevant information become available, a supplemental report will be submitted.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II COLONOVIDEOSCOPE
Common Device Name
COLONOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA   965-8520
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key23820912
Report Number9610595-2025-39849
Device Sequence Number9332012
Product Code FDF
UDI-Device Identifier04953170202315
UDI-Public04953170202315
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K100584
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source Foreign,Company Representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 12/16/2025
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberCF-Q180AL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/10/2025
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 11/18/2025
Initial Report FDA Received Date12/16/2025
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2009
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Reuse
Patient Sequence Number1
Patient SexUnknown
-
-