Model Number 1100403
Medical Device Problem Code Thermal Decomposition of Device (1071)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 11/20/2025
Type of Reportable Event Malfunction
Event or Problem Description
The manufacturer received information alleging an issue related to a simplygo device involving a thermal product malfunction.There was no report of patient harm or injury, and there was no report of medical intervention.The device was returned to the third-party service center, sieves are clogged.Once plugged in, it smells burn because the pca is burned.Inlet filter need to be replaced due to preventive intenance.All parts need to be replaced.The manufacturer's investigation is ongoing.The manufacturer's investigation is ongoing.A follow-up report will be submitted when the manufacturer's investigation is complete.
 
Additional Manufacturer Narrative
The manufacturer previously received information alleging an issue related to a simplygo device involving a thermal product malfunction.There was no report of patient harm or injury, and there was no report of medical intervention.The device was returned to the third-party service center for investigation, where the unit was inspected and error logs were checked, due to clogged sieve canisters and a burned pca, which caused a burning smell when plugged in.The inlet filter also required replacement as part of preventive maintenance.All affected components, including the pca, sieve canisters, and inlet filter, were replaced.After servicing, inspection, and testing, the device met all customer requirements, passed the functional test and quality inspection, and was packaged for return.The device was fully serviced and repaired by plexus oradea.The device completed all required testing and passed successfully.The reported failure of thermal issue is accommodated in the product risk management file as being linked to hazard with description fire, flame, smoke.Complaints for this failure are reviewed via the post market complaint trending and escalation processes to assess for the observed probability levels and compare them with the predicted levels in the risk file for the hazards linked to the failure.As of the date of submission for this report, there is no further indication that complaints for the failure in question has led to unacceptable increase in probability levels for the hazardous situations in scope, and therefore no further action will be pursued.Qa continues to monitor complaints for this issue per the cadence in the complaint trending procedures.Dhr review was performed for the complaint device serial number (b)(6), review confirms that the device met the final acceptance criteria and was released for final distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIMPLYGO
Common Device Name
GENERATOR, OXYGEN, PORTABLE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA
Manufacturer (Section G)
Respironics, Inc.
1001 murry ridge lane
murrysville PA
Manufacturer Contact
melissa rosko
1001 murry ridge lane
murrysville, PA 
+1(724)733-0200
MDR Report Key23821444
Report Number2518422-2025-112638
Device Sequence Number18513029
Product Code CAW
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111885
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Y
Reporter Type Manufacturer
Report Source Other,Foreign,Company Representative
Initial Reporter Occupation 505
Type of Report Initial,Followup
Report Date (Section B) 09/28/2026
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number1100403
Device Catalogue Number1100403
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/02/2025
Supplement Date Received by Manufacturer12/02/2025
Initial Report FDA Received Date12/16/2025
Supplement Report FDA Received Date09/28/2026
Was Device Evaluated by Manufacturer? (Y/N) No
Date Device Manufactured04/14/2023
Is the Device Labeled for Single Use? (Y/N) No
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device N
Patient Sequence Number1
Patient SexUnknown
-
-